Agilent Technologies U3606B Programmer's Reference Manual page 348

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

16
IEEE-488.2 Common Commands
*ESE
Parameter
enable_
value
[1]
Remarks
• An enable register defines which bits in the event register is reported to the
• The *CLS (clear status) command will not clear the enable register but it
• You can use the *PSC command to control whether the Standard Event
• Refer to
Example
& *ESE 16
& *ESE?
$ +16
See also
"*CLS"
"*ESR?"
"*PSC"
336
Item
Type
Numeric
Refer to
Table 16-1
for the list of bit definitions for the Standard Event register. To enable more than
one bit at a time, use the binary-weighted sum of the intended bits. For example, to set bit 2, 3, and
4, use 28 (4 + 8 + 16) as your value.
Status Byte register group. You can write to or read from an enable register.
will clear all bits in the event register.
enable register is cleared on power-on.
Figure 1-1
on page 334
on page 337
on page 340
Range of values
[1]
0 to 255
on page 13 for the status system diagram.
This command enables bit 4 (decimal value =
16) in the enable register. If an execution error
occurs, this condition will be reported to the
Status Byte register (bit 5 will be set to high).
This query returns which bits are enabled in the
register.
Default value
0
U3606B Programmer's Reference