CALibration Subsystem
Remote Calibration Procedures
22 Select the 100 Ω range. Input 100 Ω
23 Calibrate the 100 Ω point for 2-wire
24 Start the calibration.
66
Calibration steps
Resistance gain adjustment — 100 Ω
resistance to the Ω (red) and LO (black)
input terminals.
resistance measurements.
Remote commands
& CONF:RES 100
& CAL:VAL 100
& CAL?
U3606B Programmer's Reference