CONFigure Subsystem
CONFigure:CURRent:ACDC|DCAC
Example 1
& CONF:CURR:ACDC
& READ?
$ +9.535700E-02
Example 2
& CONF:CURR:ACDC 1, 0.0001
& INIT
& FETC?
$ +9.832310E-02
See also
"CONFigure?"
"INITiate[:IMMediate]"
"FETCh?"
"READ?"
98
on page 80
on page 120
on page 329
on page 331
This command configures the instrument for
AC+DC current measurements. The default
range (autorange) and resolution (5½ digits)
are used.
This command places the instrument in the
"wait-for-trigger" state, triggers a
measurement, and sends the reading to the
instrument memory and output buffer.
This command configures the instrument for
AC current measurements. The 1 A range is
μ
selected with 100
A resolution.
This command places the instrument in the
"wait-for-trigger" state, triggers a
measurement, and stores the reading in the
instrument memory.
This command transfers the reading from the
instrument memory to the output buffer.