[Sense:]Frequency:current:range:auto - Agilent Technologies U3606B Programmer's Reference Manual

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

10
SENSe Subsystem

[SENSe:]FREQuency:CURRent:RANGe:AUTO

[SENSe:]FREQuency:CURRent:RANGe:AUTO

Syntax
[SENSe:]FREQuency:CURRent:RANGe:AUTO <mode>
This command disables or enables autoranging for frequency measurements
via the current path. Autoranging is convenient because the instrument
automatically selects the range for each measurement based on the input
signal detected.
[SENSe:]FREQuency:CURRent:RANGe:AUTO?
This query returns a boolean value that represents the frequency autoranging
setting: 0 or 1
Parameter
auto
Remarks
• Autorange thresholds:
• With autoranging enabled, the instrument selects the appropriate range
• Selecting a discrete range (see
• The instrument enables autoranging after a Factory Reset (*RST
244
Item
Type
Boolean
Down range at: <10% of range
Up range at: >120% of range
based on the input signal detected.
"[SENSe:]FREQuency:CURRent:RANGe[:UPPer]"
autoranging.
command).
Range of values
0|1|OFF|ON
Default value
1
on page 242) will disable
U3606B Programmer's Reference