Agilent Technologies U3606B Programmer's Reference Manual page 147

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

Remarks
• You can allow the instrument to automatically select the measurement
• The range is set to autoranging (AUTO) when the <range> parameter is
• Autorange thresholds:
• If the input signal is greater than can be measured on the selected range
• If a range change is in progress (due to the autoranging setting) when you
U3606B Programmer's Reference
range using autoranging or you can select a fixed range using manual
ranging. Autoranging is convenient because the instrument decides which
range to use for each measurement based on the input signal. For faster
measurements, use manual ranging on each measurement (some additional
time is required for autoranging since the instrument has to make a range
selection).
omitted. The resolution is set 5½ digits (MIN) when the <resolution>
parameter is omitted.
Down range at: <10% of range
Up range at: >120% of range
(manual ranging), the instrument gives an overload indication.
Positive overload: "+9.900000E+37" or "OL" from the front panel.
Negative overload: "-9.900000E+37" or "–OL" from the front panel.
query the instrument, the value "+9.910000E+37" will be returned. Wait
for the instrument to select an appropriate range before querying the
instrument again.
MEASure Subsystem
MEASure[:VOLTage]:AC?
7
135

Advertisement

Table of Contents
loading

Table of Contents