[SENSe:]CURRent:ACDC|DCAC:RANGe:AUTO

Syntax
[SENSe:]CURRent:ACDC|DCAC:RANGe:AUTO <mode>
This command disables or enables autoranging for AC+DC current
measurements. Autoranging is convenient because the instrument
automatically selects the range for each measurement based on the input
signal detected.

[SENSe:]CURRent:ACDC|DCAC:RANGe:AUTO?

This query returns a boolean value that represents the AC+DC current
autoranging setting: 0 or 1
Parameter
auto
Remarks
• Autorange thresholds:
• With autoranging enabled, the instrument selects the appropriate range
• Selecting a discrete range (see
• The instrument enables autoranging after a Factory Reset (*RST
U3606B Programmer's Reference
Item
Type
Boolean
Down range at: <10% of range
Up range at: >120% of range
based on the input signal detected.
"[SENSe:]CURRent:ACDC|DCAC:RANGe[:UPPer]"
autoranging.
command).
[SENSe:]CURRent:ACDC|DCAC:RANGe:AUTO
Range of values
0|1|OFF|ON
SENSe Subsystem
Default value
1
on page 203) will disable
10
205