11
SOURce Subsystem

[SOURce:]SQUare[:LEVel][:IMMediate]:AMPLitude

[SOURce:]SQUare[:LEVel][:IMMediate]:AMPLitude

Syntax
[SOURce:]SQUare[:LEVel][:IMMediate]:AMPLitude <value>
This command sets the amplitude for the square-wave output.
[SOURce:]SQUare[:LEVel][:IMMediate]:AMPLitude?
This query returns a numeric value that represents the amplitude of the
square-wave output.
N O T E
Parameter
value
Example
& SQU:AMPL 10.0
& SQU:AMPL?
$ +1.000000E+01
284
This command cannot change the range of the output. You should use the
SOURce:VOLTage:RANge
Item
Type
Numeric
command to change range of the output.
Range of values
Range
Min
S1
0 V
S2
0 V
S1S2 (AUTO)
0 V
This command sets the immediate
square-wave output voltage to 10.0 V.
This query returns the square-wave output
voltage.
Default value
Max
30 V
Min
8 V
30 V