[Sense:]Continuity:range:auto - Agilent Technologies U3606B Programmer's Reference Manual

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

[SENSe:]CONTinuity:RANGe:AUTO

Syntax
[SENSe:]CONTinuity:RANGe:AUTO <mode>
This command disables or enables autoranging for continuity tests.
Autoranging is convenient because the instrument automatically selects the
range for each measurement based on the input signal detected.

[SENSe:]CONTinuity:RANGe:AUTO?

This query returns a boolean value that represents the continuity autoranging
setting: 0 or 1
Parameter
auto
Remarks
• Autorange thresholds:
• With autoranging enabled, the instrument selects the appropriate range
• Selecting a discrete range (see
• The instrument disables autoranging after a Factory Reset (*RST
U3606B Programmer's Reference
Item
Type
Boolean
Down range at: <10% of range
Up range at: >120% of range
based on the input signal detected.
page 215) will disable autoranging.
command).
[SENSe:]CONTinuity:RANGe:AUTO
Range of values
0|1|OFF|ON
"[SENSe:]CONTinuity:RANGe[:UPPer]"
SENSe Subsystem
10
Default value
1
on
217

Advertisement

Table of Contents
loading

Table of Contents