[Source:]Sense - Agilent Technologies U3606B Programmer's Reference Manual

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

[SOURce:]SENSe

Syntax
[SOURce:]SENSe {EXTernal|INTernal}
This command selects the sense signal source. To read the sense signal inputs
from the rear output terminals of the instrument, select EXTernal sense.
The valid options are:
• EXTernal: Enables external sense.
• INTernal: Disables external sense.
N O T E

[SOURce:]SENSe?

This query returns a string value that represents the sense signal source of the
instrument.
Parameter
sense
Example
& SENS EXT
& SENS?
$ EXT
U3606B Programmer's Reference
The rear output SENSE terminals (S+ and S–) are internally shorted to the
output terminals at the front panel.
Item
Type
Discrete
Range of values
EXTernal|INTernal
This command sets the signal source to read
the sense signal inputs from the rear output
terminals of the instrument.
This query returns the selected sense signal
source.
SOURce Subsystem
[SOURce:]SENSe
Default value
INTernal
257
11

Advertisement

Table of Contents
loading

Table of Contents