11
SOURce Subsystem

SOURce:VOLTage:RANGe

SOURce:VOLTage:RANGe

Syntax
SOURce:VOLTage:RANGe
This command sets the range for the voltage output. You can select to either
operate the U3606B in the S1 (30 V/1 A) range, the S2 (8 V/3 A) range, S2m
(1000 mV/3 A) range, or the S1S2 auto-range (30 V/1 A, 8 V/3 A). The range
selected is applied for all output operations (CV mode, CC mode, square-wave
output, and sweep functions).
Parameter
value
Remarks
• You cannot change the range when the output is enabled
• The over-current limit will always be set to the maximum value with respect
Example
& SOUR:VOLT:RANG 8
See also
"SOURce:CURRent:RANGe"
"SOURce:SQUare:RANGe"
260
{30 V|8 V|1 V|AUTO|MAX|MIN|DEF}
Item
Type
Numeric
(OUTPut[:STATe] ON command). The output should always be in standby
state (OUTPut[:STATe] OFF command) before the range or output
function is changed.
to the range selected.
Range of values
• 30 V|MAX|DEF
• 8 V
• 1 V|MIN
• AUTO
This command sets the voltage output range to
S2 (8 V/3 A).
on page 261
on page 262
Default value
30 V