2
CALCulate Subsystem

CALCulate:DBM:REFerence

CALCulate:DBM:REFerence

Syntax
CALCulate:DBM:REFerence <value>
This command selects the dBm reference resistance. This reference value
affects both the dBm and dB functions in the CALCulate:FUNCtion
command.
CALCulate:DBM:REFerence?
This query returns a numeric value that represents the dBm reference
resistance.
Parameter
value
[1]
Remarks
• The dBm reference resistance does not reset when calculation functions are
• The dBm reference resistance value is stored in the nonvolatile memory. It
32
Item
Type
[1]
Numeric
Integers only. All decimal parts are truncated. For example, 60.7 ohms is truncated to 60 ohms.
enabled by the CALCulate[:STATe] command, nor when the
CALCulate:FUNCtion command is sent with CALCulate:STATe set to
ON.
is not affected by a power-on cycle, Factory Reset
(*RST command), Instrument Preset (SYSTem:PRESet command), or
function change.
Range of values
1 ohm to 9999 ohms
Default value
600 ohms