Zero Offset Adjustments - Agilent Technologies U3606B Programmer's Reference Manual

Dc power supply
Hide thumbs Also See for U3606B:
Table of Contents

Advertisement

3
CALibration Subsystem
Remote Calibration Procedures

Zero offset adjustments

Each time you perform a zero offset adjustment, the instrument stores a new
set of offset correction constants for measurement functions and ranges. The
instrument will sequence through all required functions and ranges
automatically and store new zero offset calibration constants.
C A U T I O N
Be sure to allow the instrument to warm up and stabilize for 2 hours before
performing the adjustments. Follow the steps outlined below. Review the "Test
Considerations" described in the U3606B Service Guide before beginning this
test.
1 Select the DC voltage measurement.
2 Calibrate the zero point for DC voltage
3 Start the calibration.
4 Select the 2-wire resistance measurement.
5 Calibrate the zero point for 2-wire resistance
6 Start the calibration.
56
Never turn off the instrument during zero offset adjustment. This may
cause ALL calibration memory to be lost.
Calibration steps
Zero offset adjustment — DC voltage (short)
Connect a shorting plug between the V (red)
and LO (black) input terminals.
measurements.
Zero offset adjustment — 2-wire resistance (short)
Leave the shorting plug between the Ω (red)
and LO (black) input terminals connected.
measurements.
Remote commands
& CONF:VOLT:DC
& CAL:VAL 0
& CAL?
& CONF:RES
& CAL:VAL 0
& CAL?
U3606B Programmer's Reference