Analog Input
Configuration
Thermo Fisher Scientific
The Analog Input Configuration menu displays a list of the 8 analog input
channels available for configuration. This screen is only displayed if the I/O
expansion board option is installed. Configuration includes entering the
Descriptor, Units, Decimal Places, choice of 1–10 points in the table, and
corresponding number of points selected. This menu is not displayed if the
Output Conc Cal option is selected.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config.

Descriptor

The Descriptor screen allows the user to enter the descriptor, or name, for
the selected analog input channel. The descriptor is used in datalogging
and streaming data to report what data is being sent out. The descriptor
may be from 1 to 3 characters in length, and defaults to IN1 to IN8 (user
input channel number).
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > Descriptor.
ANALOG INPUT CONFIG:
>CHANNEL 1
CHANNEL 2
CHANNEL 3
CHANNEL 4
CHANNEL 5
RANGE
AVG
DIAGS ALARM
ANALOG INPUT 01 CONFIG:
>DESCRIPTOR
UNITS
DECIMAL PLACES
TABLE POINTS
POINT 1
RANGE
AVG
DIAGS ALARM
POINT 2
ANALOG INPUT DESCRIPTOR:
CURRENTLY: IN1
IN1
ABCDEFGHIJKLMN
OPQRSTUVWXYZ
0123456789 ./-
RANGE
AVG
DIAGS ALARM
Instrument Controls Menu
IN1
IN2
IN3
IN4
IN5
IN1
V
2
2
BKSP
PAGE
SAVE
Model 42i Instruction Manual
Operation
3-53