Thermo Scientific 42i Instruction Manual page 344

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Geysitech (Bayern-Hessen) Protocol
Geysitech Commands
Measurements reported in
response to DA command
Single Range Mode
Dual/Auto Range Mode
Operating and Error
D-6
Model 42i Instruction Manual
Example:
Geysitech Protocol with transmission of three concentrations (Instrument
ID is 1, Operation Status is 03, Error Status is 04):
Data Query String:
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>0000000000 <SP>002 <SP>
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681
0000000000<SP><CR>
The attached concentrations are in the selected gas units. The
measurements that are attached to the response if not valid in a particular
mode then a value of 0.0 will be reported.
The following measurements reported in response to DA command are for
the Model 42i.
The 3 measurements reported in single range mode include:
NO
NO
2
NO
x
The 6 measurements reported in dual or auto range modes include:
low NO
low NO
low NO
high NO
high NO
high NO
See
Table D–1
Model 42i.
Status
<STX>DA<CR>
Address First Concentration(E-format)=25.78
2
x
2
x
for operating status and
Address+2
Third Concentration=11.75
Table D–2
for error status for the
Thermo Fisher Scientific
Address+1

Advertisement

Table of Contents
loading

Table of Contents