Figure 7-11. Replacing The Solenoid Valve - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Ozonator Assembly
Replacement
Thermo Fisher Scientific
6. Re-install the measurement bench. Refer to
Measurement Bench and Lowering the Partition
Solenoid Valve
Mounting Clip
Figure 7–11. Replacing the Solenoid Valve
Use the following procedure to replace the ozonator assembly
12).
Equipment Required:
Ozonator assembly
Wrench, 5/8-inch
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
1. Refer to
"Removing the Measurement Bench and Lowering the
Partition
Panel" in this chapter to lower the partition panel, then
proceed to the next step below.
Servicing
Ozonator Assembly Replacement
"Removing the
Panel" in this chapter.
(Figure 7–
Model 42i Instruction Manual
7-21