Set Defaults

Factory Calibrate Oven
Thermistor

Low and High Points

Thermo Fisher Scientific
The Set Defaults screen is used to reset the configuration values to factory
defaults.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Gas Therm > Set Defaults.
RESTORE DEFAULT CAL:
FOR PERM OVEN HTR THERM
RANGE
AVG
DIAGS ALARM
The Factory Calibrate Oven Thermistor menu is used for calibrating the
permeation oven heater thermistor to either low point, high point, or set
defaults.
In the Main Menu, choose Service > Perm Oven Settings > Factory
Cal Oven Therm.
The Calibrate Oven Thermistor Low Point screen is used to view and set
the permeation oven thermistor resistance. The low and high point screens
function the same way.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Oven Therm > Low or High Point.
CAL GAS THERM LOW POINT:
CURRENTLY:
0= 3850 Ohms
SET TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
RESTORE
FACTORY CAL OVEN THERM:
>LOW POINT
HIGH POINT
SET DEFAULTS
RANGE
AVG
DIAGS ALARM
04000
Ohms
SAVE
RESTORE DEFAULT CAL:
FOR PERM OVEN GAS THERM
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
RANGE
AVG
DIAGS ALARM
Model 42i Instruction Manual
Operation
Service Menu
RESTORE
TO?x
3-89