C-Link Protocol Commands
Measurement Configuration
B-38
Model 42i Instruction Manual
Send:
set span
Receive:
set span ok
gas unit
This command reports the current gas units (ppb, ppm, μg/m
The example reports that the gas unit is set to ppb.
Send:
gas unit
Receive:
gas unit ppb
set gas unit unit
unit = | ppb | ppm | μg/m
This command sets the gas units to ppb, ppm, μg/m
example below sets the gas units to mg/m
Send:
set gas unit mg/m3
Receive:
set gas unit mg/m3 ok
meas mode
This command reports which measurement mode (NO/NO
NO
) is active. The example below reports that the measurement mode is
x
set to NO.
Send:
meas mode
Receive:
meas mode no
set meas mode mode
mode = | no/nox | no | nox | pre |
This command sets the instrument to NO/NO
mode, or manual NO
x
the manual NO mode.
Send:
set meas mode no
Receive:
set meas mode no ok
pres comp
This command reports whether pressure compensation is on or off. The
example below shows that pressure compensation is on.
Send:
pres comp
Receive:
pres comp on
set pres comp onoff
These commands turn the pressure compensation on or off. The example
below turns pressure compensation off.
Send:
set pres comp off
Receive:
set pres comp off ok
3
3
| mg/m
|
3
.
mode. The example below sets the instrument to
3
, or mg/m
3
3
, or mg/m
. The
, NO, or
x
(auto) mode, manual NO
x
Thermo Fisher Scientific