Servicing
Analog Output Testing
7-28
Model 42i Instruction Manual
5. Check that the meter is displaying the zero value. If the meter reading
differs by more than one percent, the analog outputs should be
adjusted. Refer to the
follows.
6. Press
to set the output to full-scale.
The Output Set To line displays Full-Scale.
7. Check that the meter is displaying a full-scale value. If the meter
reading differs by more than one percent, the analog outputs should be
adjusted. Refer to the
8. Press
to reset the analog outputs to normal.
Table 7–4. Analog Output Channels and Rear Panel Pin Connections
Voltage
Pin
Channel
1
14
2
33
3
15
4
34
5
17
6
36
Ground
16, 18, 19, 35, 37
Table 7–5. Analog Input Channels and Rear Panel Pin Connections
Input Channel
Pin
1
1
2
2
3
3
4
5
5
6
6
7
7
9
8
10
Ground
4, 8, 11, 14
"Analog Output
Calibration" procedure that
"Analog Output
Calibration" procedure b.
Current
Pin
Channel
1
15
2
17
3
19
4
21
5
23
6
25
Current Output Return
13, 16, 18, 20, 22, 24