Serial Communication
Parameters
Thermo Fisher Scientific
Appendix D
Geysitech (Bayern-Hessen) Protocol
This appendix provides a description of the Geysitech (Bayern-Hessen or
BH) Protocol Interface and is supported both over RS-232/485 as well as
TCP/IP over Ethernet.
The Geysitech Commands that are implemented are explained in detail in
this document. The Geysitech protocol support for the iSeries enables the
user to perform the functions of reading the various concentrations and to
trigger the instrument to be in sample/zero/span mode if valid for that
instrument. This is achieved by using the supported Geysitech commands
listed below.
For details of the Model 42i Geysitech Protocol specification, see the
following topics:
"Serial Communication
"TCP Communication
"Instrument
Address" on page
"Abbreviations
Used" on page
"Basic Command
Structure" on page
"Block
Checksum" on page
"Geysitech
Commands" on page
The following are the communication parameters that are used to configure
the serial port of the iSeries to support Geysitech protocol.
Number of Data bits
Number of Stop bits
Parity
Data rate
Parameters" on page
D-1
Parameters" on page
D-2
D-2
D-2
D-2
D-3
D-2
: 7 or 8
: 1 or 2
: None, Odd, or Even
: 1200 to 115200 Baud (9600 is default)
Model 42i Instruction Manual
D-1