C-Link Protocol Commands
Calibration
B-32
Model 42i Instruction Manual
set zero dur value
This command sets the zero duration to value, where value represents the
zero duration in minutes. The following example sets the zero duration to
15 minutes.
Send:
set zero duration 15
Receive:
set zero duration 15 ok
zs period
This command reports the zero/span (z/s) period. The following example
reports that the zero/span period is 24 hours.
Send:
zs period
Receive:
zs period 24 hr
set zs period value
This command sets the zero/span period to value, where value represents
the zero/span period in hours. The following example sets the zero/span
period to 24 hours.
Send:
set zs period 24
Receive:
set zs period 24 hr ok
zs avg time
This command reports the zero/span (z/s) averaging time in seconds. The
following example reports that the zero/span averaging time is 30 seconds,
according to
Table
B–2.
Send:
zs avg time
Receive:
zs avg time 5:30 sec
set zs avg time selection
This command sets the zero/span averaging time, according to
The following example sets the zero/span averaging time to 120 seconds.
Send:
set zs avg time 8
Receive:
set zs avg time 8 ok
Table
B–2.