Figure 3-7. Pin-Out Of Rear Connector In Auto Range Mode; Table 3-4. Default Analog Outputs In Auto Range Mode - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Operation
Range Menu
3-12
Model 42i Instruction Manual
By default, in the auto range mode, the analog outputs are arranged on the
rear panel connector as shown in
and pin connections. Auto range mode may be selected from the
Mode
Select" on page 3-76.
Figure 3–7. Pin-Out of Rear Connector in Auto Range Mode
Table 3–4. Default Analog Outputs in Auto Range Mode
Channel
Connector Pin
1
14
2
33
3
15
4
34
5
17
6
36
Ground
16, 18, 19, 35, 37
Note All channels are user definable. If any customization has been made
to the analog output configuration, the default selections may not apply.
Figure
3–7. See
Table 3–4
I/O Terminal Pin
Description
1
NO Analog Output
3
NO
5
NO
7
Range Status:
half-scale = high range
zero scale = low range
9
None
11
None
2, 4, 6, 8, 10, 12
Signal Ground
for channels
"Range
Analog Output
2
Analog Output
x
Thermo Fisher Scientific