Converter
Temperature
Min and Max Converter
Temperature Limits
Permeation Oven
Gas Temperature
Thermo Fisher Scientific
COOLER TEMPERATURE:
ACTUAL MIN
SET MIN TO:
RANGE
AVG
The Converter Temperature screen displays the current converter
temperature and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 300 to 700 °C. The actual alarm setpoints should
be set for the installed converter. If the converter temperature reading goes
beyond either the minimum or maximum limit, an alarm is activated and
the alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu.
In the Main Menu, choose Alarms > Converter Temp.

CONVERTER TEMPERATURE:

ACTUAL
>MIN
MAX
RANGE
AVG
The Minimum Converter Temperature alarm limit screen is used to change
the minimum converter temperature alarm limit. The minimum and
maximum converter temperature screens function the same way.
In the Main Menu, choose Alarms > Converter Temp > Min or Max.
CONVERTER TEMPERATURE:
ACTUAL MIN
SET MIN TO:
RANGE
AVG
The Permeation Oven Gas Temperature screen displays the current
permeation oven gas temperature and sets the minimum and maximum
alarm limits. Acceptable alarm limits range from 25.00 to 50.00 °C. If the
o
-20.0
C
o
-10.0
C ?

INC/DEC
SAVE VALUE
DIAGS ALARM
o
320.7
C
o
300.0
C
o
350.0
C
DIAGS ALARM
o
300.0
C
o
310.0
C ?

INC/DEC
SAVE VALUE
DIAGS ALARM
Operation
Alarms Menu
Model 42i Instruction Manual
3-69