C-Link Protocol Commands
Communications Configuration
B-48
Model 42i Instruction Manual
The following example sets the instrument to accept the "set mode local"
or "set mode remote" commands.
Send:
set allow mode cmd 1
Receive:
set allow mode cmd 1 ok
Table B–10. Allow Mode Command Values
Allow Mode Command
Value
0
Ignore (default)
1
Accept
power up mode
This command reports the current power up mode setting either 0 =
local/unlocked or 1 = remote/locked according to
value is 0; power up in local/unlocked mode. The following example shows
that the instrument is configured to power up in the remote/locked mode.
Send:
power up mode
Receive:
power up mode 1
set power up mode value
This command is used to configure the instrument to power up in the
local/unlocked mode (value = 0) or the remote/locked mode (value = 1),
according to
Table
B–11.
If the instrument is set to power up in the local/unlocked mode, the keypad
can be used to make changes via the front panel. If the instrument is set to
power up in the remote/locked mode, changes can not be made from the
front panel. The example that follows sets the instrument to power up in
remote/locked mode.
Send:
set power up mode 1
Receive:
set power up mode 1 ok
Table B–11. Power Up Mode Values
Power Up Mode Command
Value
0
Local/Unlocked (default)
1
Remote/Locked Mode
program no
This command reports the analyzer's model information and program
version number, which will be dependant on the current version.
Table
B–11. The default
Thermo Fisher Scientific