Figure 7-16. Replacing The Pressure Transducer - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Servicing
Pressure Transducer Calibration
Pressure Transducer
Calibration
7-32
Model 42i Instruction Manual
2. Disconnect plumbing from the pressure transducer assembly. Note the
plumbing connections to facilitate reconnection.
3. Disconnect the pressure transducer cable.
Figure 7–16. Replacing the Pressure Transducer
4. Remove the two pressure transducer assembly retaining screws and
remove the pressure transducer assembly.
5. To install the pressure transducer assembly, follow previous steps in
reverse.
6. Calibrate the pressure transducer. Refer to the
Calibration" procedure that follows.
Use the following procedure to calibrate the pressure transducer.
Equipment Required:
Vacuum pump
WARNING The service procedures in this manual are restricted to qualified
service representatives.
Retaining Screws (2)
Pressure Transducer Assembly
"Pressure Transducer
Thermo Fisher Scientific