Table C-3. Write Coils For 42I - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Thermo Fisher Scientific
Register
Variable
Number
121 & 122
EXT ALARMS
*
The analog inputs 1–8 will not show if the Ouput Conc Cal option is set to YES in the Instrument
Configuration screen.
**
BKG and COEF items are displayed only if Ouput Conc Cal option is set to YES in the Instrument
Configuration screen.
IMPORTANT NOTE The addresses in the following tables are Protocol
Data Unit (PDU) addresses. Verify the coil number on your MODBUS
master to ensure that it matches the coil number on the instrument.
Note Writing 1 to the coil number shown in the following table will
initiate the "action triggered" listed in the table. This state must be held for
at least 1 second to ensure the instrument detects the change and triggers
the appropriate action.
Note The coils within each coil group in the following table are mutually
exclusive and will not be triggered if there is conflict. Before you assert (1)
one coil in a group, make sure the other coils in the group are de-asserted
(0).
Table C–3. Write Coils for 42 i
Coil Number
Action Triggered
100
NOT USED
101
ZERO MODE
102
SPAN MODE
103
NO MODE
104
NOX MODE
105
NOT USED
106
NOT USED
107
SET BACKGROUND
108
CAL TO LO SPAN
109
AOUTS TO ZERO
110
AOUTS TO FS
111
CAL TO HI SPAN
112
NOT USED
113
NOT USED
Coil Group
Zero/Span Mode
Zero/Span Mode
Measure Mode
Measure Mode
Analog Out Test
Analog Out Test
MODBUS Protocol
MODBUS Parameters Supported
Zero/Span Option
Zero/Span Option
I/O Expansion Board Option
I/O Expansion Board Option
Model 42i Instruction Manual