Communication Protocol; Streaming Data Configuration - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Operation
Instrument Controls Menu

Streaming Data Configuration

Streaming Data Interval
3-38
Model 42i Instruction Manual

COMMUNICATION PROTOCOL:

CURRENTLY:
SET TO:
RANGE
AVG
The Streaming Data Configuration menu is used to allow for configuration
of the 8 streaming data output items, streaming interval, current data
format, and current timestamp setting. The Choose Stream Data submenu
displays a list of the analog output signal group choices to choose from.
Choices are Concentrations, Other Measurements, and Analog Inputs (if
the I/O expansion board option is installed).
Note The selected item list is different depending on the
AUTO/MANUAL MODE and RANGE MODE settings. Changing
either of these settings will produce a different set of streaming data items.
If either of these modes is going to be changed on a regular basis, then the
user must configure each set of data separately after changing the modes to
each setting.
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config.
STREAMING DATA CONFIG
>INTERVAL
ADD LABELS
PREPEND TIMESTAMP
ADD FLAGS
ITEM 1
RANGE
AVG
The Streaming Data Interval screen is used to adjust how frequently a new
record will be generated. The following interval times are available: 1, 2, 5,
10, 20, 30, 60, 90, 120, 180, 240, and 300 seconds.
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Interval.
CLINK
STREAMING ?

CHANGE VALUE
SAVE VALUE
DIAGS ALARM
10 SEC
NO
YES
YES
NO
DIAGS ALARM