C-Link Protocol Commands
Communications Configuration
B-44
Model 42i Instruction Manual
addr nm
This command reports the IP netmask.
Send:
addr nm
Receive:
addr nm 255.255.255.0
set addr nm address
This command sets the nm address, where address consists of four numbers
ranging from 0-255 inclusive, separated by ".".
Note This command cannot be used when DHCP is on. Refer to the
DHCP command that follows for additional information.
Send:
set addr nm 255.255.255.0
Receive:
set addr nm 255.255.255.0 ok
addr ntp
This command reports the IP address for the network time protocol server.
See "Network Time Protocol Servier" in the "Communications Settings"
section of the "Operation" chapter for more information.
Send:
addr ip
Receive:
addr ip 192.168.1.15
set addr ip address
This command sets the NTP time server address, where address consists of
four numbers ranging from 0-255 inclusive, separated by ".".
Send:
set addr ip 192.168.1.15
Receive:
set addr ip 192.168.1.15 ok
baud
This command reports the current baud rate for the serial port
(RS232/RS485). The example below reports that the current baud rate is
9600 baud.
Send:
baud
Receive:
baud 9600
set baud rate
rate = | 1200 | 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200 |
This command sets the instrument baud rate. The example below sets the
instrument's baud rate to 9600.
Note After the command is sent, the baud rate of the sending device must
be changed to agree with the instrument.
Thermo Fisher Scientific