Operation
Instrument Controls Menu

Other Measurements

Analog Inputs
3-40
Model 42i Instruction Manual
The Other Measurements screen allows the user to assign one of the other
available measurement types to the selected streaming data item. The
currently selected item is shown by "<--" after it. Once an item is selected,
pressing
will save the selected streaming data item. BKG and COEF
items are displayed only if Ouput Conc Cal option is set to YES in the
Instrument Configuration screen.
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Select Item > Other
Measurements.
OTHER MEASUREMENTS:
>NONE
INT TEMP
CHAMBER TEMP
COOLER TEMP
NO2 CNV TEMP
RANGE
AVG
CHAMBER PRES
FLOW
PMT VOLTS
NO BKG
NOx BKG
LO NO COEF
LO NO2 COEF
LO NOx COEF
HI NO COEF
HI NO2 COEF
HI NOx COEF
OZONATOR FLOW
EXT ALARMS
The Analog Inputs screen allows the user to assign an analog input signal
(none or analog inputs 1–8) to the selected streaming data item. The
currently selected item is shown by "<--" after it. Once an item is selected,
pressing
will save the selected streaming data item. The analog
inputs 1–8 will not show if the Ouput Conc Cal option is set to YES in the
Instrument Configuration screen.
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Select Item > Analog Inputs.
DIAGS ALARM