LCD Module
Replacement
Thermo Fisher Scientific
LCD Module Screws (2)
Front Panel Board
Mounting Studs (2)
Ribbon Cables (3)
Bottom Support
Figure 7–22. Replacing the Front Panel Board and the LCD Module
Use the following procedure to replace the LCD module
Equipment Required:
LCD module
Philips screwdriver
CAUTION If the LCD panel breaks, do not to let the liquid crystal contact
your skin or clothes. If the liquid crystal contacts your skin or clothes, wash
it off immediately using soap and water.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
Do not remove the LCD panel or frame from the LCD module.
The LCD polarizing plate is very fragile, handle it carefully.
LCD Module
▲
LCD Module Replacement
Retaining Screw
Two-Wire Connector
(Figure
7–22).
▲
▲
▲
Model 42i Instruction Manual
Servicing
7-45