C-Link Protocol Commands
Calibration
B-28
Model 42i Instruction Manual
These commands report NO, NO
mode, or the high and low range coefficients in dual or auto range mode. If
the mode is incorrect, the instrument responds with "can't, wrong
settings". The example below reports that the NO coefficient is 1.000.
Send:
no coef
Receive:
no coef 1.000
set no coef value
set no2 coef value
set nox coef value
set high no coef value
set high no2 coef value
set high nox coef value
set low no coef value
set low no2 coef value
set low nox coef value
These commands set the NO, NO
values to value, where value is a floating-point representation of the
coefficient. The example below sets the NO coefficient to 1.005.
Send:
set no coef 1.005
Receive:
set no coef 1.005 ok
no gas
no2 gas
nox gas
high no gas
high no2 gas
high nox gas
low no gas
low no2 gas
low nox gas
These commands report NO, NO
to auto-calibrate NO, NO
commands are only available in dual and auto range mode. If the mode is
incorrect, the instrument responds with "can't, wrong settings". The
example below reports that the NO low span gas concentration is 240.0
ppb.
Send:
low no gas
Receive:
low no gas 2400E-1 ppb
set no gas value
set no2 gas value
set nox gas value
set high no gas value
, and NO
coefficients in single range
2
x
, and NO
coefficients to user-defined
2
x
, and NO
span gas concentrations used
2
x
, and NO
coefficients. The high and low
2
x