Operation
Service Menu

Permeation Oven Selection

Factory Calibrate Gas Thermistor

Low and High Points

3-88
Model 42i Instruction Manual
CAL OVEN THERM (RESISTOR):
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
The Permeation Oven Setpoint screen is used to select the permeation oven
state to not present, or the desired temperature choice of 30, 35, and 45 °C.
In the Main Menu, choose Service > Perm Oven Settings > Perm Oven
Selection.
PERM OVEN SETPOINT:
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
The Factory Calibrate Gas Thermistor menu is used for calibrating the
permeation oven gas thermistor to either low point, high point, or set
defaults.
In the Main Menu, choose Service > Perm Oven Settings > Factory
Cal Gas Therm.
FACTORY CAL GAS THERM:
>LOW POINT
HIGH POINT

SET DEFAULTS

RANGE
AVG
The Calibrate Gas Thermistor Low Point screen is used to view and set the
permeation oven thermistor resistance. The low and high point screens
function the same way.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Gas Therm > Low or High Point.
3850 Ohms
04000
Ohms

MOVE CURSOR
SAVE
DIAGS ALARM
NOT PRESENT
o
45
C ?

MOVE CURSOR
SAVE
DIAGS ALARM
DIAGS ALARM
Thermo Fisher Scientific