Specifications
A.8
USB
Table A-13. USB Memory Device Specifications
Characteristics
USB compatibility
Memory file format
Number of files
USB size
Table A-14. Data Transfer Characteristics
Ventilator data description
Trends capacity
Events capacity
Monitorings capacity
A.9
Pneumatic
Table A-15. Airway Resistances
Inspiratory
1.0 mbar at 30 lpm flow ± 0.1 mbar
3.7 mbar at 60 lpm flow ± 0.1 mbar
Table A-16. Patient Circuit Resistances
Adult Double Limb
2 mbar at 60 lpm flow
a. Includes exhalation valve
b. Values obtained from the manufacturer's directions for use.
Table A-17. Air Inlet Resistance (Filter)
1.1 cmH
O (1.079 mbar) at 30 lpm flow
2
Table A-18. Oxygen Inlet Specifications
Maximum
pressure
50 kPa (7 psi)
Table A-19. Performance Specifications
Working
pressure
5 mbar –
55 mbar
A-8
Exhalation
0.5 mbar at 30 lpm ± 0.1 mbar
1.1 mbar at 60 lpm ± 0.1 mbar
a
Pediatric Double Limb
b
2 mbar at 30 lpm flow
Maximum
flow
15 lpm
Sound pressure
Maximum
level
pressure limit
30 dBA (per NF
60 mbar
EN ISO 17510-1
test conditions)
Supported formats
USB flash memory USB 2.0 or USB 1.1
USB 32 bit format (sector size: 512 - 2,048 bytes)
Maximum 999
128 MB to 4 GB
Capacity
86 MB
512 KB or 5,500 events
42 MB / 48 hours
Internal
compliance
(ventilator)
.0001 l/mbar
Inspiratory
triggering
response time
(Ttr)
100 ms
User's Manual