Rem Alarm Indicator - Covidien Force FX-C User Manual

Electrosurgical generator
Table of Contents

Advertisement

REM Alarm Indicator

REM Alarm Indicator
This indicator illuminates red until you properly apply a REM Polyhesive patient return
electrode to the patient and connect it to the generator. Then the indicator illuminates
green. (When you connect an electrode without the REM safety feature, the indicator
does not illuminate.)
If the REM system senses an alarm condition, the indicator flashes red until you correct
the alarm condition—then the indicator illuminates green. (If you are using a return
electrode without the REM safety feature, the red indicator light is extinguished when
you correct the alarm condition.)