Class I Equipment (Iec 60601-1); Type Cf Equipment (Iec 60601-1)/Defibrillator Proof; Liquid Spillage (Iec 60601-2-2:2006 Clause 44.3 And Iec 60601-2-2:2009 Clause 201.11.6.3); Voltage Transients (Emergency System Mains Transfer) - Covidien Valleylab LS10 Service Manual

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

Class I Equipment (IEC 60601-1)

Accessible conductive parts cannot become live in the event of a basic insulation failure
due to the way in which they are connected to the protective earth conductor.
Type CF Equipment (IEC 60601-1)/Defibrillator
Proof
This generator provides a high degree of protection against electric shock,
particularly regarding allowable leakage currents. It is type CF isolated (floating)
output and may be used for procedures involving the heart.
This generator complies with IEC 60601-1:1988 + A1:1991 + A2:1995 and IEC 60601-
1:2005 specifications for "defibrillator proof" designation and IEC 60601-2-2:2006 and
IEC 60601-2-2:2009.
Liquid Spillage (IEC 60601-2-2:2006 Clause 44.3
and IEC 60601-2-2:2009 Clause 201.11.6.3)
The generator is constructed so that liquid spillage in normal use does not wet electrical
insulation or other components which when wetted are likely to adversely affect the
safety of the equipment.
Voltage Transients (Emergency System Mains
Transfer)
The generator continues to operate normally with no errors or system failures when
transfer is made between line AC and an emergency system-voltage source. (IEC 60601-
1:1988 + A1:1991 + A2:1995 clause 49, IEC 60601-1:2005 clause 11.8, IEC 60601-2-
2:2006 clause 51.101, and IEC 60601-2-2:2009 clause 201.11.8)
Electromagnetic Compatibility (IEC 60601-1-2 and

IEC 60601-2-2)

The generator complies with the appropriate IEC 60601-1-2 and 60601-2-2
specifications regarding electromagnetic compatibility.
Notice
The generator requires special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in the generator service manual.
Portable and mobile RF communications equipment can affect the generator. Refer to the EMC
information provided in the Valleylab LS10, LS Series Single Channel Vessel Sealing Generator
Service Manual.