USB Port

The software provides an asynchronous serial-communications interface for
communicating with an externally-connected device.
USB B-type connector, enumerated with a serial/COM port with 115200 bps baud, 8 data
bits, 1 stop bit, no flow control setting in the laptop.

Potential Equalization Conductor Terminal

A potential equalization conductor terminal is provided to allow connection of the
generator to the ground.

Radio Frequency Identification (RFID)

The RFID module is located above the LigaSure port. The intended use of the RFID module
is to identify the inserted LigaSure instrument and configure the generator with the data
included in the RFID tag.
Frequency Range: 13.56 MHz
RF Output Power: 68.17 dBuV/m @ 3 meters
Type of Antenna: Integral Loop Antenna
Modulation: Amplitude-shift Keying (ASK)
Mode of Operation (Simplex/Duplex): Duplex
Contains Transmitter Module FCC ID: 2AAVI-JDK1901
Contains IC ID: 11355A-JDK1901
Low Frequency (50/60 Hz) Leakage Current (IEC 60601-2-2)
Enclosure source
current, ground open
Source current, patient
leads, all outputs
Sink current at high
line, all inputs
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator
< 300 μA
Normal polarity, intact ground: < 10 μA
Normal polarity, ground open: < 50 μA
Reverse polarity, ground open: < 50 μA
Mains voltage on applied part: < 50 μA
< 50 μA
Performance Characteristics