Operating Parameters
Ambient temperature
range
Relative humidity
Atmospheric pressure
Warm-up time
Transport and Storage
Ambient-temperature
range
Relative humidity
Atmospheric pressure
Duration of storage
Duty Cycle
Under maximum-output settings and rated-load conditions (30 ohm load) the generator
is suitable for activation times 5 seconds on, 15 seconds off, for one hour. With lesser
settings and loads, you can activate the generator for greater durations without
generating excessive internal temperatures.
Internal Battery
Battery for Real Time
Clock
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
+10° C to +40° C (50° F to 104° F)
30% to 75% non-condensing
700 millibars to 1060 millibars
If transported or stored at temperatures outside the
operating temperature range, allow one hour for the
system to reach room temperature before use.
-30° C to +65° C (-22° F to 149° F)
25% to 85% (non-condensing)
500 millibars to 1060 millibars
If the energy platform is stored for over one year
complete a periodic safety check. Contact Covidien
Service for information.
Battery type – 3 V lithium button cell
Battery life – 5 years
Performance Characteristics
2-3