28.4.4. Set tare value

Syntax: UT_TARE CR LF, where TARE – tare value
Possible answers:
UT_OK CR LF
UT_I CR LF
ES CR LF
Notice:
This protocole uses the dot character as a decimal point for separating
the decimal fraction part.

28.4.5. Send the stable result in basic unit

Syntax: S CR LF
Possible answers:
S_A CR LF
S_E CR LF
S_I CR LF
S_A CR LF
MASS FRAME
Frame format:
1
2-3
S
space
stability
Example:
S CR LF – computer command
S _ A CR LF - command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value
in basic unit is returned.
- command executed
- command comprehended but cannot be executed
- command not recognised (possible wrong tare format)
- command accepted and in progress
- time limit for stable result exceeded
- command comprehended but cannot be executed
- command accepted and in progress
- mass value in basic unit is returned
4
5
space
6
7-15
16
sign
mass
space
- 158 -
17
18
19
20
unit
CR
21
LF