Laser Voltage Is Out Of Specification; Source Voltage Is Out Of Specification; Power Supply Voltages Are Out Of Tolerance; Humidity Is Too High - Thermo Scientific iS5N Series User Manual

Table of Contents

Advertisement

4
Troubleshooting
Error Messages

Laser Voltage Is Out of Specification

Source Voltage Is Out of Specification

Power Supply Voltages Are Out of Tolerance

Humidity Is Too High

Laser Temperature Is Out of Specification

Interferometer Temperature Is Out of Specification

62
Nicolet iS5N Spectrometer User Guide
1. Remove any accessories or samples from the sample compartment.
2. Choose Experiment Setup > Diagnostic > Align and wait for spectrometer alignment to
complete.
3. Contact Technical Support for assistance; see
1. Make sure the source holder is properly seated in the instrument.
2. Replace the source. See
3. Contact Technical Support for assistance; see
1. Choose Experiment Setup > Diagnostic and click the Power icon. Check that the
current value for +12 Volt is within specifications. If it is out of the specified range,
replace the power supply.
2. If other values on the Power diagnostic screen are out of their specified ranges, contact
Technical Support for assistance; see
1. If the instrument is equipped with an optional purge kit, check that the purge gas flow is
set properly; see
Setting Purge Gas Controls
2. Check the humidity indicator and replace the desiccant and indicator; see
Desiccant
for instructions.
1. Make sure the room temperature is between 15 °C and 35 °C (between 60 °F and 95 °F),
and allow the instrument adequate time to stabilize.
2. Contact Technical Support for assistance; see
1. Make sure the room temperature is between 15 °C and 35 °C (between 60 °F and 95 °F),
and allow the instrument adequate time to stabilize.
2. Contact Technical Support for assistance; see
Contacting
Replacing the
Source.
Contacting
Contacting
for details.
Contacting
Contacting
Us.
Us.
Us.
Us.
Us.
Replacing the