Connecting AC Power

Fabius MRI can be operated at mains voltages
from 100 V to 240 V. For 240 V operation, use a
special cable (see CAUTION statement below).
1 Push power plug into supply mains socket.
2 Switch on the machine by using the system
power switch on the rear of the machine.
WARNING
Only connect device to approved "hospital
grade"outlets having the appropriate ground-
ing reliablity.
WARNING
Do not connect life-supporting devices to the
auxiliary power outlet(s) of the anesthesia
machine; if the mains/hospital power supply
fails, devices connected to the auxiliary power
outlet(s) will not be supplied by the anesthesia
machine's battery backup and will discon-
tinue operation.
WARNING
Be careful in handling the power cord and
main power plug. These parts still contain
minor magnetic components. The power cord
can be attracted to MRI system.
CAUTION
Fabius MRI is intended to be used on 120 V /
60 Hz supplies in the USA. The corresponding
power cable is listed in the list of accessories.
Connection of Fabius MRI to a 240 V / 60 Hz sup-
ply is only allowed if it is a center-tapped-single-
phase circuit and the power cable is a certified,
hospital-grade cable 10 A/ 250 V AC of max. 3 m
(9.8 ft) length with:
– a NEMA 6-15 HG plug,
– a SJT 3*18 AWG cable and
– a C13 connector.
In this case Dräger recommends the mains cable
with part no. 1868152.