Operation; Controlled Mandatory Ventilation (Cmv) - Dräger MIcroVent Operating Instructions Manual

Ventilator
Table of Contents

Advertisement

Operation

Operation
WARNING !
Always use ventilator that has been cleaned
and disinfected and has been successfully
tested to be ready for operation.

Controlled Mandatory Ventilation (CMV)

For ventilator rates of 5 to 60 bpm.
NOTE: During CMV, the MicroVent ventilator limits the
minimum ventilator rate to 5 bpm.
1
Set switch for ventilator modes to CMV
2
Set knobs »f« and »V
«.
T
3
Set knob »T
«.
insp.
4
Set knob »P
« initially to 60 cmH
max
5
Set knob »PEEP« initially to 0 cmH
6
Set O
concentration on blender.
2
7
Set alarm limits for minute volume.
8
On/Off switch to »ON«.
Display:
Self-test
Software xx.xx
When the patient has been connected:
Check display of minute ventilation and re-adjust
for the specific patient.
WARNING !
Observe pressure gauge in order to identify im-
proper ventilation. Operator must always be
ready to intervene if alarm indicates a situation
dangerous to the patient.
30
O.
2
O.
2
Controlled Mandatory
Ventilation (CMV)
2
2 3 1
4
5
7
8
D
6