Equipotential Bonding

e.g. intra-cranial and intra-cardiac operations.
Use cable ref. 83 01 349.
1 Connect the terminal on the back to an equipo-
tential bonding point in the operating room.
Connect one end of the earth cable to one of the
connecting pins located on the back of the
Fabius OS.
Connect the other end of the earth cable to the
specified equipotential bonding point, e.g., on
the operating table or ceiling lamp.
Connect equipotential bonding to the auxiliary
equipment.

Connecting AC Power

Fabius OS can be operated at mains voltages from
100 V to 240 V.
Push power plug into supply mains socket.
2 Switch on the machine by using the ON/OFF
switch on the rear of the machine.
WARNING
Only connect device to approved "hospital
grade" outlets having the appropriate ground-
ing reliablity.
WARNING
Do not connect life-supporting devices to the
auxiliary power outlet(s) of the anesthesia
workstation; if the mains/hospital power sup-
ply fails, devices connected to the auxiliary
power outlet(s) (optional) will not be supplied
by the anesthesia workstation's battery
backup and will discontinue operation.