Using Cmv - Dräger Oxylog 2000 Operating Instructions Manual

Emergency & transport ventilator
Hide thumbs Also See for Oxylog 2000:
Table of Contents

Advertisement

WARNING !
Always use ventilator that has been cleaned
and disinfected and has been successfully
tested to be ready for operation.
"Care", page 40
"Preparation", page 14
"Checks of Readiness for Operation", page 23

Using CMV

CMV = Controlled Mandatory Ventilation
For ventilator rates of 5 to 40 bpm.
NOTE: The Oxylog 2000 ventilator limits the minimum
ventilator rate to 5 bpm during CMV.
1
Set switch for ventilator modes to CMV
NOTE: For quick pre-setting you may set knobs »Rate«
and »V
« to matching colors on dial scale. This will
T
adjust ventilation parameters to settings for infant,
pediatric, and adult patients, respectively.
2 Set »Rate« and »V
« knobs.
T
Body weight
Rate
(kg)
(bpm)
Green range:
30 to 40
infants
(7.5 to 20)
Blue range:
20 to 30
children
(20 to 40)
Brown range:
5 to 20
adults
(over 40)
V
T
(L)
0.1 to 0.3
0.3 to 0.8
0.8 to 1.5
Operation
Using CMV
2
2
1
29