Displaying Settings And Measured Values - Dräger Oxylog 2000 Instructions For Use Manual

Emergency ventilator
Hide thumbs Also See for Oxylog 2000:
Table of Contents

Advertisement

When spontaneous breathing is insufficient:
change to SIMV or IPPV/SIPPV.
The automatic »Paw low« warning for a disconnection
and the »Leak« warning are not effective in CPAP
mode.
Check that connections are tight, and with mask
ventilation check that the mask is correctly fitted.

Displaying settings and measured values

The ventilation mode and expiratory minute volume MV
are continuously shown in the top line of the display.
Specific ventilation parameters appear in the bottom line
of the display, depending on the ventilation mode set
(examples):
For IPPV/SIPPV
Inspiration time
Inspiratory flow
Mean pressure
End expiratory pressure
Peak pressure
Respiration rate
(SIPPV only)
For SIMV
Inspiratory flow
Mean pressure
End expiratory pressure
Peak pressure
Respiration rate
Tidal volume
For CPAP
Pos. airway pressure
Respiration rate
Tidal volume
In the event of an alarm, the parameters displayed in the
bottom line will be "overwritten" by the warning messages.
Displaying measured values:
M
1 Briefly press the »
Info« key.
*
The criterion for PEEP is the display on the pressure gauge.
The electronic display can deviate, as the measuring point is not
identical (see description of operating principles on page 48).
Tinsp = 2.0 s
FLOW = 30.0 L/min
MEAN = 15.0 mbar
PEEP* = 5.0 mbar
PEAK = 20.0 mbar
Frequency = 10 1/min
FLOW = 30.0 L/min
MEAN = 15.0 mbar
PEEP* = 5.0 mbar
PEAK = 20.0 mbar
Frequency = 10 1/min
V
exp. = 450 mL
T
CPAP* = 8.0 mbar
Frequency = 10 1/min
V
exp. = 450 mL
T
Displaying settings and measured values
1
D
Operation
Using CPAP
Oxylog 2000
13