Adjustments; Test Software And Information Available On The Web; Tme Downloads; Agilent Fieldfox Customer Support Web Pages - Agilent Technologies FieldFox N9923AN Service Manual

Rf vector network analyzer
Table of Contents

Advertisement

Tests and Adjustments
RF Vector Network Analyzer

Test Software and Information Available on the Web

N9923AN

Adjustments

Frequency Accuracy
Locks to Ext Ref 10 MHz signal via DAC. Stores DAC setting to use when not in External
Reference mode.
Phase Lock Accelerator
This process creates factors for various phase lock capture speeds used during
measurements. No test equipment is required.
Error Terms
This adjustment process updates the correction information used by CalReady (i.e. the
Factory Cal). It uses an ECal to perform a one port cal on each port, then completes a two
port cal using the thru and loads.
Note: The Error Terms adjustment should be performed every twelve months to preserve
calibration accuracy of CalReady.
Test Software and Information Available on the Web

TME Downloads

Information and download links for the N7841A application are available at
www.agilent.com/find/calibrationsoftware. This web site provides links for downloading
software and user instructions. Note that the N7800A software must be installed prior to
installing the N7841A and it requires a separate license.
The web page for the N7841A download also provides a link to the N7841A help. This help
system provides user instructions and lists of required equipment. It may be accessed prior
to downloading the software.

Agilent FieldFox Customer Support Web Pages

The following URL provides updated help and additional information for using TME with
FieldFox. Consult it for the most up-to-date information.
http://na.tm.agilent.com/fieldfox/tme_info.html
3-8
Service Guide N9923-90005