200 Hz Procedure; Khz Procedure - Agilent Technologies PSA Series Manual

Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

ESA Functional Tests
EMI Detector Test
Figure 4-4 Equipment Setup

200 Hz Procedure

1. Connect the equipment as shown in Figure 4-3. on page 159.
For the E7401A, it is not necessary to connect the AMPTD REF OUT to the
NOTE
analyzer's 50Ω input. It switches internally.
2. Press
on the analyzer and wait for the preset routine to finish. Set the analyzer
Preset
up by pressing the following keys:
,
Input/Output
Amptd Ref Out (f=50 MHz)
,
FREQUENCY
Center Freq
,
,
Span
2
kHz
,
AMPLITUDE
Attenuation
,
,
BW/Avg
Res BW
, 25,
AMPLITUDE
—dBm
, 20,
AMPLITUDE
—dBm
Peak Search
,
MEASURE
Meas at Marker
3. Wait for the Measure at Marker routine to finish and then record the peak, quasi-peak,
and average values measured by the analyzer. When the measurement is complete, a
box will appear in the upper left hand corner of the screen with the most recent
measurement information. Record these on your worksheet (in
in 200 Hz Procedure.

9 kHz Procedure

1. Connect the equipment as shown in Figure 4-3. on page 159.
NOTE
For the HP E7401A, it is not necessary to connect the AMPTD REF OUT to
the analyzer's 50Ω input. It switches internally.
160
(On)
, 50,
MHz
, 5,
,
,
dB
More
Y Axis Units
,
, 300,
200 Hz
Video BW
Hz
(E7401A only)
(E7402A, E7403A, E7404A, and E7405A only)
,
dBm
Table 4-2 on page
162)
Chapter 4