5 Reader Parameters
Default/Reset/Save
Understanding and controlling your reader's active, saved, and default settings is critical to its successful
operation.
5-7-13 Resets
Resets ("A" commands) affect only the current settings (active memory) and are not saved for power-on.
5-7-14 Saved for Power-On
Power-on parameters ("Z" commands) are saved to NOVRAM and recalled and loaded into current
parameters when power is cycled to the reader or the <Arp> command is issued.
5-7-15 Defaults
Defaults are Omron Microscan firmware settings or saved customer settings that can be recalled, either
by software or hardware reset.
5-7-16 Customer Default Parameters
Customer default parameters (saved by <Zc>) are the same set of parameters as power-on parameters
but are saved in a different, isolated section of NOVRAM. This allows a user essentially to create a
backup set of parameters that can be recalled in the event that the current parameters or power-on
parameters have been accidentally changed or are no longer desired.
It is important to note that a hardware default does not affect customer default parameters. For exam-
ple, a user that has inadvertently changed communication settings and saved them with a <Z> com-
mand, may not know the correct settings or have the capability to communicate within those settings.
By first doing a hardware default to restore the known
recall the previous customer saved settings with an <Arc> or <Zrc> command.
5-7-17 Omron Microscan Default Parameters
Omron Microscan default parameters are contained in the firmware and cannot be changed.
Software Defaults
Default parameters can be recalled (loaded into current settings) with <Ard> command or recalled and
saved for power-on with the <Zrd> command.
5-70
Reset
Reset and Recall Microscan Defaults
Reset and Recall Power-On Parameters
Reset and Recall Customer Default Parameters
Save Current Settings for Power-On
Save Current Settings as Customer Default Parameters for Power-On
Recall Microscan Default Parameters and Save for Power-On
Recall Customer Default Parameters and Save for Power-On
Function
Omron
Microscan defaults, the user can then
V460-H Industrial Handheld DPM Reader User Manual
Serial Cmd
<A>
<Ard>
<Arp>
<Arc>
<Z>
<Zc>
<Zrd>
<Zrc>