Agilent Technologies 1290 Infinity User Manual page 99

Lc injectors htc/hts
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

Wash Station Reference Points
Wash1/Wash2 Reference Points
For the DLW Wash Station Option, the reference positions for Wash1 and
Wash2 are the two holes on top of the Wash Station assembly, see
page 99.
The lower needle guide should be centered in these holes with the bottom of
the lower needle guide lightly touching the surface of the Wash Station
assembly.
Figure 49
1 Press "HOME" on the handheld controller and select the sequence
2 Enter Wash1 and adjust the X-, Y-, Z-positions until the lower needle guide
3 Repeat steps 1 – 3 for Wash2 by using the path: Menu/Setup/Objects/Wash
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
DLW Wash Station Wash1/Wash2 Reference Point
Menu/Setup/Objects/Wash Stations/Wash1 .
of the Injection Unit centers on top of the Wash station assembly. Fine tune
the Z-position until the bottom of the lower needle guide is flush with the
top surface.
Stations/Wash2
LC Injector HTC/HTS Description and Installation
3
Object Positions
Figure 49