Page 3
In This Guide... In This Guide... This manual covers the Agilent 1290 Infinity Thermostatted Column Compartment (TCC) (G1316C). 1 Introduction to the Column Compartment This chapter gives an introduction to the TCC and an instrument overview. 2 Site Requirements and Specifications This chapter provides information on environmental requirements, physical and performance specifications.
Page 4
10 Identifying Cables This chapter summarizes information on all cables. 11 Hardware Information This chapter describes the TCC in more detail on hardware and electronics. 12 Appendix This chapter provides addition information on safety, legal and web. 1290 Infinity TCC User Manual...
4 How to optimize the Column Compartment Optimizing the Column Compartment Using Additional Heater Devices Delay Volume and Extra-Column Volume How to Configure the Optimum Delay Volume How to Achieve Higher Injection Volumes How to Achieve High Throughput 1290 Infinity TCC User Manual...
Page 6
Replacing Head Parts of Column Switching Valve Replacing Valve Heads Preparing the Column Compartment for Transportation Replacing the Module Firmware 9 Parts and Materials for Maintenance Plastic Parts Leak Parts Heater and Cooling Devices Valve Options Overview Accessory Kits 1290 Infinity TCC User Manual...
Page 7
Setting the 8-bit Configuration Switch (without On-board) LAN Early Maintenance Feedback Instrument Layout 12 Appendix General Safety Information The Waste Electrical and Electronic Equipment (WEEE) Directive (2002-96-EC) Lithium Batteries Information Radio Interference Sound Emission Agilent Technologies on Internet 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Introduction to the Column Compartment Overview of the Module System Overview Column-Identification System Column Switching Valve (Optional) Typical Applications Bio-inert Materials This chapter gives an introduction to the TCC and an instrument overview. Agilent Technologies...
The Agilent 1290 Infinity Thermostatted Column Compartment is a stackable temperature-controlled column compartment for LC. It is available as standalone module or as a component of an Agilent 1290 Infinity system. It is used for heating and cooling to meet extreme requirements of retention time reproducibility.
This depends on several factors: flow rate, setpoint temperature, ambient temperature and column dimensions. The higher the flow rate, the faster the column equilibrates (due to thermostatted mobile phase). 1290 Infinity TCC User Manual...
Page 12
The equilibration of the column may take longer. Stability of the pressure signal is a good indication for equilibrium. Figure 1 Equilibration of Heat Exchanger and Column Temperature The temperature calibration and verification is described in the Service Manual. 1290 Infinity TCC User Manual...
It allows to read and write column-specific information to and from an optionally available column-identification tag (part of the HPLC-starter-kits available for 1220, 1260 and 1290 Infinity pumps or separately as PN 5062-8588). Figure 2 Column-Identification System Table 1 on page 14 shows the information that can be stored.
Page 14
(and vice versa). If no column switching valve is installed, both sides are updated at the same time. If an 8 Position/9 Port valve is installed, no tag is updated for avoiding ambiguity as more than 2 columns might be installed. 1290 Infinity TCC User Manual...
Introduction to the Column Compartment Column Switching Valve (Optional) Column Switching Valve (Optional) Figure 3 Location of Column Switching Valve 1290 Infinity TCC User Manual...
Page 16
Before switching the valve, switch off the pump or set the flow to zero. Keeping the flow on N O T E while the valve is switched can cause exceeding the maximum pressure. This will stop method or sequence execution. Figure 4 Column 1 Active Figure 5 Column 2 Active 1290 Infinity TCC User Manual...
Page 17
The sample is injected into series-connected precolumn and analytical column. After the valve has switched, the analytical column flow continues in normal direction. Only the precolumn is back-flushed, eluting highly retained peaks directly to the detector. Figure 6 Precolumn Back-flushing 1290 Infinity TCC User Manual...
Page 18
Sample Enrichment and Sample Cleanup 2pos/6port valve (G4231A/B) and 2pos/10port valve (G4232A/B/C) Advantages: • Easy automation of sample preparation • Higher reproducibility • Increased productivity and sensitivity Sample Enrichment 1290 Infinity TCC User Manual...
Page 19
After the valve switches, an additional pump backflushes the matrix components out of the pre-column to waste, while the analytes are separated on the main column. Backflushing prepares the pre-column for the next injection. 1290 Infinity TCC User Manual...
Page 20
Core of the alternating column regeneration is the Agilent 1200 Infinity Series 2 position / 10 port valves, which allows simultaneous analysis on one column while a second identical column is flushed and equilibrated. Figure 7 Alternate Column Regeneration (Time Scheme) 1290 Infinity TCC User Manual...
Bio-inert Materials Bio-inert Materials For the Agilent 1260 Infinity Bio-inert LC system, Agilent Technologies uses highest quality materials in the flow path (also referred to as wetted parts), which are widely accepted by life scientists, as they are known for optimum inertness to biological samples and ensure best compatibility with common samples and solvents over a wide pH range.
1290 Infinity TCC User Manual Site Requirements and Specifications Site Requirements Physical Specifications Performance Specifications Valve Specifications Extended Specifications on G1316C Solvent Information This chapter provides information on environmental requirements, physical and performance specifications. Agilent Technologies...
In case of emergency it must be possible to disconnect the instrument from the power line at any time. ➔ Make sure the power connector of the instrument can be easily reached and unplugged. ➔ Provide sufficient space behind the power socket of the instrument to unplug the cable. 1290 Infinity TCC User Manual...
Page 25
Never operate your instrumentation from a power outlet that has no ground connection. ➔ Never use a power cord other than the Agilent Technologies power cord designed for your region. Use of unsupplied cables WA R N I N G Using cables not supplied by Agilent Technologies can lead to damage of the electronic components or personal injury.
Page 26
Do not store, ship or use your module under conditions where temperature fluctuations could cause condensation within the module. ➔ If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation. 1290 Infinity TCC User Manual...
Operating altitude Up to 2000 m (6562 ft) Non-operating altitude Up to 4600 m (15091 ft) For storing the module Safety standards: Installation category II, Pollution degree 2 For indoor use only. IEC, CSA, UL 1290 Infinity TCC User Manual...
Housing All materials recyclable. All specifications are valid for distilled water at ambient temperature (25 °C), set point at N O T E 40 °C and a flow range from 0.2–5 ml/min. 1290 Infinity TCC User Manual...
Kit contains 1x 8 Pos. / 9 port, 1200 barr and 1 x 8 Pos. / 9 port , 600 bar valve head Table 7 G4231A Agilent 1290 Series, 2 Position / 6 Port valve head, 600bar Type Specification Liquid contacts...
Page 30
Stator - 0.38 mm (0.015 in), Rotor Seal- 0.30 mm (0.012 in) Port to Port Volume 0.51 µL Maximum pressure 1200 bar Table 9 G4232A Agilent 1290 Series, 2 Position / 10 Port micro valve head , 600bar Type Specification Liquid contacts PEEK, Stainless Steel Port size...
Page 31
Site Requirements and Specifications Valve Specifications Table 11 G4232C Agilent 1290 Series, 2 Position / 10 Port valve head , 600bar Type Specification Liquid contacts PEEK, Stainless Steel Port size Accepts 10-32 male threaded fittings Flow passage diameters Stator - 0.15 mm (0.006 in), Rotor Seal- 0.20 mm (0.008 in) Port to Port Volume 0.16 µL...
Page 32
PEEK, Ceramic Port size Accepts 10 – 32 male threaded fittings Flow passage diameters 0.38 mm (0.015 in) Stator 0.30 mm (0.012 in) Rotor Seal Port to Port Volume 1.71 µL Maximum pressure 600 bar 1290 Infinity TCC User Manual...
Page 33
Liquid contacts PEEK, Ceramic Port size Accepts 10 – 32 male threaded fittings Flow passage diameters 0.25 mm (0.010 in) Port to Port Volume 0.51 µL (pre.column volume) 1.33 µL (post.column volume) Maximum pressure 600 bar 1290 Infinity TCC User Manual...
Extended Specifications on G1316C Extended Specifications on G1316C The 1290 Infinity Thermostatted Column Compartment (G1316C) is usable from 10 °C below ambient up to 80 °C for flow ranges up to 5 ml/min and up to 100 °C for flow ranges up to 2.5 ml/min. Additional heating devices are...
• Chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, di-isopropylether) such ethers should be filtered through dry aluminium oxide which adsorbs the peroxides. 1290 Infinity TCC User Manual...
Page 36
• Mixtures of carbon tetrachloride with 2-propanol or THF. Solvent information for parts of the 1260 Infinity Bio-inert LC system For the Agilent 1260 Infinity Bio-inert LC system, Agilent Technologies uses highest quality materials (see “Bio-inert Materials”...
Page 37
(a mixture of concentrated hydrochloric and nitric acid). Zirconium Oxide Zirconium Oxide (ZrO ) is inert to almost all common acids, bases and solvents. There are no documented incompatibilities for HPLC applications. 1290 Infinity TCC User Manual...
Page 38
Most data available refers to room temperature (typically 20 – 25 °C, 68 – 77 °F). If corrosion is possible, it usually increases at higher temperatures. If in doubt, consult additional resources. 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Installing the Module Unpacking the module Damage to the module Delivery Checklist Optimizing the Stack Configuration One Stack Configuration Two Stack Configuration Installing the Column Compartment Installing Valve Heads Installation of Heater Devices Flow Connections of the Column Compartment...
Damage to the module If the delivery packaging shows signs of external damage, please call your Agilent Technologies sales and service office immediately. Inform your service representative that the instrument may have been damaged during shipment. "Defective on arrival" problems C A U T I O N If there are signs of damage, please do not attempt to install the module.
Optimizing the Stack Configuration Optimizing the Stack Configuration If your column compartment is part of an Agilent 1290 Infinity system, you can ensure optimum performance by installing the following configuration. This configuration optimizes the system flow path, ensuring minimum delay volume.
Figure 10 on page 43). This configuration optimizes the flow path for minimum delay volume and minimizes the bench space required. The Agilent 1290 Infinity Binary Pump should always be installed at the bottom of the stack. Figure 9 Recommended stack configuration for 1290 Infinity with binary pump (front view)
Page 43
Installing the Module Optimizing the Stack Configuration Figure 10 Recommended stack configuration 1290 Infinity with binary pump (rear view) 1290 Infinity TCC User Manual...
A slightly longer capillary is required between the pump and autosampler. (See Figure 11 on page 44 and Figure 12 page 45). Figure 11 Recommended two stack configuration for 1290 Infinity with binary pump (front view) 1290 Infinity TCC User Manual...
Page 45
Installing the Module Optimizing the Stack Configuration Figure 12 Recommended two stack configuration for 1290 Infinity with binary pump (rear view) 1290 Infinity TCC User Manual...
Selection of valve port positions can fail, if the instrument does not know the properties of the installed valve. Always switch off the instrument when replacing the valve head. 1290 Infinity TCC User Manual...
Page 47
1 Remove the 5 screws, which hold the lock in position. 2 Remove the dummy valve head by unscrewing the cap nut and removing it from the valve drive. 3 Place the module on the bench in a horizontal position. 1290 Infinity TCC User Manual...
Page 48
6 Connect the power cable to the power connector at the rear of the module. The security lever will prevent that the cover is opened while the power cord is connected to the module. 1290 Infinity TCC User Manual...
Page 49
When the line power button stands out and the green light is off, the module is turned off. N O T E The module was shipped with default configuration settings. For changing these settings, N O T E refer to section Setting the 8-bit configuration switch. 1290 Infinity TCC User Manual...
Installing the Valve Heads The valve drives are factory-installed in the 1290 Infinity Thermostatted Column Compartment , in the 1290 Infinity Flexible Cube, and in the 1290 Infinity Universal Valve Drive. The valve heads are interchangeable and can be easily mounted.
Page 51
When unscrewing the transportation lock (TCC only), To remove the valve dummy, loosen the nut manually. push it back until the last screw is removed - the valve rail is spring-loaded 1290 Infinity TCC User Manual...
Page 52
For bio-inert applications, always use dedicated bio-inert parts, which can be identified by the bio-inert symbol or other markers described in this manual. ➔ Do not mix bio-inert and non-inert modules or parts in a bio-inert system. 1290 Infinity TCC User Manual...
Page 53
To have the valve correctly recognized by the Agilent Infinity valve drive you must have the N O T E valve drive powered off for at least 10 s. 1290 Infinity TCC User Manual...
Page 54
While turning the valve you should feel some resistance from the drive. Otherwise the pins on the valve shaft are not properly fitted into the corresponding adapters on the valve drive. 1290 Infinity TCC User Manual...
Page 55
Push the valve head until it snaps in and stays in the rear Power on or power-cycle your module, so the valve head position. (TCC only) gets recognized during module initialization. N O T E Power Off the Infinity valve drive for at least 10 s. 1290 Infinity TCC User Manual...
Depending on the application, these heater devices can be fixed at various places. N O T E Information about the usage of these heater devices can be found in the Agilent 1290 Infinity LC System - System Manual and Quick Reference (G4220-90301).
Page 57
For the G1316C, an additional heat exchanger can be installed on the Carrier for heat exchanger (G1316-89200) using 3 screws ( Screw 2.5 mm hex (0515-1052), included to part number for carrier) as shown in Figure 16 page 57. Figure 16 Installing the low dispersion heat exchangers 1290 Infinity TCC User Manual...
Page 58
Some examples are shown in Figure 17 on page 58. Figure 17 Arrangements of the bio-inert solvent heating devices in the 1290 Series TCC (G1316C) If the solvent heating devices are installed in the central position on the TCC’s heat N O T E...
Page 59
Strong force/torque will damage SST cladded PEEK capillaries C A U T I O N Be careful when installing stainless steel cladded PEEK capillaries. ➔ Follow the instructions given in the technical note Installation of Stainless Steel Cladded PEEK Capillaries. (G5611-90120). 1290 Infinity TCC User Manual...
Do not operate the instrument in an explosive atmosphere. Press release buttons and remove front cover to gain Place the column on the left heat exchanger assembly access to heater area. and connect the capillaries to the column. 1290 Infinity TCC User Manual...
Page 61
Route tubings from modules above through the openings system, or if an Agilent autosampler is located on top, in the funnel holder (top) and the plastic bottom part. connect the corrugated tubing to the waste outlet. Remove small plastic plugs first. 1290 Infinity TCC User Manual...
Page 62
The installation of the column compartment has now been completed. Always operate the TCC with the front cover in place for proper thermostatting conditions N O T E and to protect the column area against strong drafts from the ouside. 1290 Infinity TCC User Manual...
Page 63
➔ The volume of substances should be reduced to the minimum required for the analysis. ➔ Do not operate the instrument in an explosive atmosphere. 1290 Infinity TCC User Manual...
Page 64
“Column-Identification System” on page 13. N O T E The internal volumes of the heat exchanger assemblies comprise a volume of 3 µl (left) and 6 µl (right). The internal capillary diameter is 0.17 mm. 1290 Infinity TCC User Manual...
Fix the column with the column clip from the accessory If the column compartment is not part of an Agilent LC kit. system, or if an Agilent autosampler is located on top, connect the corrugated tubing to the waste outlet. 1290 Infinity TCC User Manual...
The installation of the column compartment has now been completed. Always operate the TCC with the front cover in place for proper thermostatting conditions N O T E and to protect the column area against strong drafts from the outside. 1290 Infinity TCC User Manual...
January 2013), please follow procedure as described in the Appendix of the Agilent 1260 Infinity Bio-inert Quaternary LC System Manual. 1 Slide the fitting on the capillary. Let the capillary jut out 5 mm. 1290 Infinity TCC User Manual...
Page 68
Installing the Module Installing Capillaries 2 Insert the fitting to the receiving port and push the capillary to the bottom of the port. 3 Finger tighten the nut into the port until snug. 1290 Infinity TCC User Manual...
Page 69
Installing the Module Installing Capillaries 4 Use Fitting mounting tool (5043-0915) or a 5 mm hex wrench for fixing the fitting (maximum torque 0.8 Nm). 1290 Infinity TCC User Manual...
N O T E left or right heat exchanger, see Figure 19 on page 70 and Figure 20 on page 71. The Agilent logo should always be at front. Figure 19 Column-Identification Tag for Left Heat Exchanger 1290 Infinity TCC User Manual...
Page 71
Figure 20 Column-Identification Tag for Right Heat Exchanger Column Clip For better positioning of the column on the heat exchanger a column clip is available (see “Accessory Kits” on page 155). Figure 21 Column clip 1290 Infinity TCC User Manual...
Page 72
Installing the Module Placing Columns 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual How to optimize the Column Compartment Optimizing the Column Compartment Using Additional Heater Devices Delay Volume and Extra-Column Volume Delay Volume Extra-Columwn Volume How to Configure the Optimum Delay Volume How to Achieve Higher Injection Volumes...
• Keep the left and right heat exchanger temperature the same unless you do specific applications. • Assure that the front cover is always closed. 1290 Infinity TCC User Manual...
Using Additional Heater Devices Using Additional Heater Devices The optimization, installation, interconnection, and specific settings when using additional heating devices are described in the Agilent 1290 Infinity LC System - System Manual and Quick Reference (G4220-90301). 1290 Infinity TCC User Manual...
The delay volume in a system includes the volume in the pump from the point of mixing, connections between pump and autosampler, volume of the flow path through the autosampler and connections between autosampler and column. 1290 Infinity TCC User Manual...
The extra-column volume is minimized with the Agilent 1290 Infinity/Agilent 1260 Infinity LC system due to the narrow-bore (0.12 mm i.d.) tubing, the low-volume heat exchangers in the column compartment and the Max-Light cartridge cell in the detector.
0.12 mm i.d. capillaries. The built-in 3 μl and 6 μl heat exchangers are for backwards compatibility and should only be used in case a conventional method needs to be run on the system even though in this case also the low dispersion heat exchanger could be used. 1290 Infinity TCC User Manual...
“Sample Enrichment” page 79. The valve can be conveniently located in the thermostatted column compartment or in the flexible cube. Sample Enrichment 1290 Infinity TCC User Manual...
The two columns then alternate in this way for the whole sequence of injections. The second pump is only required to flush an isocratic mixture through the column and so can be a simpler pump than the 1290 Infinity pumps. For instance a 1200 Series isocratic pump would be sufficient to perform this task.
1290 Infinity TCC User Manual Troubleshooting and Diagnostics Overview of the Module’s Indicators and Test Functions Status indicators Power Supply Indicator Module Status Indicator Available Tests vs User Interfaces Agilent Lab Advisor Software Overview about the troubleshooting and diagnostic features.
Normally, temperature calibration is not required throughout the lifetime of the instrument. However, in order to comply with local regulatory requirements, calibration and verification may be required. The following sections describe these functions in detail. 1290 Infinity TCC User Manual...
Figure 22 Location of Status indicators Power Supply Indicator The power supply indicator is integrated into the main power switch. When the indicator is illuminated (green) the power is ON. 1290 Infinity TCC User Manual...
In such a case try to re-boot the module or try a cold-start (see “Special Settings” on page 193. Then try a firmware update (see “Replacing the Module Firmware” on page 143). If this does not help, a main board replacement is required. 1290 Infinity TCC User Manual...
• Preferred tool should be the Agilent Lab Advisor software, see “Agilent Lab Advisor Software” on page 86. • The Agilent ChemStation B.04.02 and above may not include any maintenance/test functions. • Screenshots used within these procedures are based on the Agilent Lab Advisor software. 1290 Infinity TCC User Manual...
For details refer to the Agilent Lab Advisor software help files. The Instrument Utilities is a basic version of the Lab Advisor with limited functionality required for installation, use and maintenance. No advanced repair, troubleshooting and monitoring functionality is included. 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Error Information What Are Error Messages General Error Messages Timeout Shutdown Remote Timeout Lost CAN Partner Leak Leak Sensor Open Leak Sensor Short TCC Error Messages Compensation Sensor Open Compensation Sensor Short Left Fan Failed...
In all cases, error propagation is done via the CAN bus or via an APG remote cable (see documentation for the APG interface). 1290 Infinity TCC User Manual...
Check the vacuum degasser for an error The degasser failed to generate sufficient condition. Refer to the Service Manual for the vacuum for solvent degassing. degasser or the 1260 pump that has the degasser built-in. 1290 Infinity TCC User Manual...
• Ensure all CAN cables are installed correctly. Exchange the CAN cable. Defective CAN cable. Switch off the system. Restart the system, and Defective main board in another module. determine which module or modules are not recognized by the system. 1290 Infinity TCC User Manual...
Please contact your Agilent service Leak sensor not connected to the main representative. board. Please contact your Agilent service Defective leak sensor. representative. Please contact your Agilent service Leak sensor incorrectly routed, being representative. pinched by a metal component. 1290 Infinity TCC User Manual...
Suggested actions Please contact your Agilent service Defective leak sensor. representative. • Please contact your Agilent service Leak sensor incorrectly routed, being representative. pinched by a metal component. • Please contact your Agilent service representative. 1290 Infinity TCC User Manual...
If the resistance across the sensor falls below the lower limit, the error message is generated. Probable cause Suggested actions Please contact your Agilent service Defective main board. representative. 1290 Infinity TCC User Manual...
If the fan speed falls below 2 revolutions/second for longer than 5 s, the error message is generated. Probable cause Suggested actions Please contact your Agilent service Fan cable disconnected. representative. Please contact your Agilent service Defective fan. representative. Please contact your Agilent service Defective TCC board. representative. 1290 Infinity TCC User Manual...
OFF the peltier elements after a short delay, and the error message is generated. Probable cause Suggested actions Please contact your Agilent service The column compartment was switched on representative. with the top cover and foam removed. 1290 Infinity TCC User Manual...
The temperature of the right heat exchanger did not reach the temperature setpoint within the timeout threshold. Probable cause Suggested actions Please contact your Agilent service Defective right heater assembly. representative. Please contact your Agilent service Defective TCC board. representative. 1290 Infinity TCC User Manual...
Please contact your Agilent service Flex board not connected (only if all left or representative. right sensor error messages appear simultaneously). Please contact your Agilent service Defective heater assembly. representative. Please contact your Agilent service Defective TCC board. representative. 1290 Infinity TCC User Manual...
105 °C, and the error message is generated. Probable cause Suggested actions Please contact your Agilent service representative. Defective heater assembly. Please contact your Agilent service representative. Defective TCC board. 1290 Infinity TCC User Manual...
The processor checks the function of the heater circuits continually. If a defect is detected in the control circuit, the processor switches OFF the heater (peltier) assemblies, and the error message is generated. Probable cause Suggested actions Please contact your Agilent service representative. Defective TCC board. 1290 Infinity TCC User Manual...
RFID tag does not contain valid information. The RFID can be damaged if the instrument is power cycled during a write access of the tag. Probable cause Suggested actions Replace valve head, see “Valve Options RFID tag contains invalid information. Overview” on page 153. 1290 Infinity TCC User Manual...
Valve head not installed correctly. “Installing the Valve Head and Connecting Capillaries” on page 52. Replace valve head, see “Replacing Valve RFID tag defective. Heads” on page 136 Please contact your Agilent service Valve tag reader is damaged. representative. 1290 Infinity TCC User Manual...
Page 102
Error Information TCC Error Messages 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Test Functions and Calibrations Thermostat Function Test Pressure Test Temperature Calibration Temperature Calibration Procedure Calibration Problems Installing the Temperature Sensor This chapter describes the TCC’s built in test functions. Agilent Technologies...
The time required to reach 30 °C is a measure of heating efficiency. Thermostat Function Test Result A typical Thermostat Function Test profile is shown in Figure 23 on page 104. Figure 23 Typical Thermostat Function Test Profile 1290 Infinity TCC User Manual...
Page 105
During the cooling phase, the Peltier elements should cool at a rate of >2 °C/minute. During the heating phase, the temperature change should be >3 °C/minute. Defective thermostat components may cause cooling or heating rates to fall outside these limits. 1290 Infinity TCC User Manual...
Do not use the test for modules having a lower maximum pressure than the pump as this will damage the valve. For example do not use 400 bar valve in a TCC or Flex Cube in combination with a 600 bar pump. 1290 Infinity TCC User Manual...
The column thermostat is calibrated correctly when the measured temperature (using the external measuring device, “Temperature Calibration Procedure” on page 108) and the cross-over temperature (36 °C) of both heat exchangers (left and right) are within a range of ± 0.5 °C. 1290 Infinity TCC User Manual...
For the measuring and calibration process Agilent Technologies recommends a N O T E thermometer with 0.1 °C precision. Contact the local Agilent Technologies support representative for ordering information. The figures in this procedure refer to a specific type of temperature sensor (Heraeus, N O T E Quat340, quartz surface-temperature measurement sensor).
Installation of the temperature sensor is required for the temperature calibration and temperature verification procedures. The figures below refer to a specific type of temperature sensor (Heraeus, Quat340, quartz N O T E surface-temperature measurement sensor). Other sensors may require a different fixing. 1290 Infinity TCC User Manual...
Page 110
Test Functions and Calibrations Temperature Calibration Remove the front cover. Install the temperature sensor at the measurement position on the left heat exchanger. Route the sensor wire through the slit in the leak tray. Re-install the front cover. 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Maintenance Introduction to Maintenance and Repair Cautions and Warnings Overview of Maintenance Cleaning the Column Compartment Changing Column Identification Tags Adding Heater Devices Installing the capillaries Correcting Leaks Replacing Head Parts of Column Switching Valve...
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the cover is opened. ➔ Do not remove the cover of the module. ➔ Only certified persons are authorized to carry out repairs inside the module. 1290 Infinity TCC User Manual...
Page 114
For bio-inert applications, always use dedicated bio-inert parts, which can be identified by the bio-inert symbol or other markers described in this manual. ➔ Do not mix bio-inert and non-inert modules or parts in a bio-inert system. 1290 Infinity TCC User Manual...
If the valve performance shows page 136 indication of leakage or wear “Preparing the Column If the TCC shall be transported Compartment for Transportation” on page 140 “Replacing the Module If required Firmware” on page 143 1290 Infinity TCC User Manual...
Liquid dripping into the electronic compartment of your module can cause shock WA R N I N G hazard and damage the module ➔ Do not use an excessively damp cloth during cleaning. ➔ Drain all solvent lines before opening any connections in the flow path. 1290 Infinity TCC User Manual...
2 The tag needs to be placed differently, depending on whether the column is installed at the left or right heat exchanger, see Figure 26 on page 118 and Figure 27 on page 118. The Agilent logo should always be at front. 1290 Infinity TCC User Manual...
Page 118
Column-Identification Tag for Right Heat Exchanger 3 For columns with small diameter, a cable tie wrap should be used to fix the column identification tag to the column. Assure that the tie wrap does not block the front cover. 1290 Infinity TCC User Manual...
Adding Heater Devices Adding Heater Devices The additional heater devices can be arranged in the G1316C in various locations depending on the application needs. Some examples are shown below. Figure 28 Arrangements of Heater and Cooling Devices 1290 Infinity TCC User Manual...
Page 120
If the column identification system is required, fix the heater devices in the upper or lower locations or fix them right/left of the current location. The heater devices are mounted on a carrier that can be fitted to the left and/or right heat exchangers. 1290 Infinity TCC User Manual...
Page 121
( Column clip set, 8 colors (5042-9918)) for more convenient installation; mark the nuts of the capillaries attached to the columns as well with color code tags (small rings) to easily follow the flow paths in your system. 1290 Infinity TCC User Manual...
Page 122
For the G1316C, additional heat exchanger devices can be installed on the carrier Carrier for heat exchanger (G1316-89200) using 3 screws ( Screw 2.5 mm hex (0515-1052), included to part number for carrier) as shown in figure below. Figure 30 Installing the low dispersion heat exchangers 1290 Infinity TCC User Manual...
Page 123
Do not overtighten the fitting at installation. Finger tighten the nut into the port until snug. Then use the wrench and apply about 1/4 of a revolution. (Torque ~ 1.0 – 1.2 Nm) Figure 31 Fitting type depends on column type 1290 Infinity TCC User Manual...
G4232B 2pos/10port valve Preparations Identify the required capillaries in your capillary kits (“Capillary kit PN 5067-4646” on page 158) Use outmost care to avoid any void volumes caused by poor connections. N O T E 1290 Infinity TCC User Manual...
Page 125
6 Starting from position one through six (ten, respectively), fasten the fittings on the heat exchanger. 7 Fasten all fittings on attached modules (autosampler, detector, additional pumps). Fit all unused valve ports with a plastic plug. 8 Push the valves into the rear positions. 1290 Infinity TCC User Manual...
Page 126
9 Place the capillaries that go to another module or waste into the capillary guides to prevent squeezing them when closing the front cover. 10 Stow any excess lengths of the capillaries. 11 Perform a final leak-check. 1290 Infinity TCC User Manual...
Page 127
Pos. 2: Connection between Ports 1-2, 3-4, 5-6 active Column 2 = right Example shows setup with flow directed 1290 typical from bottom to top. Flow direction from top to bottom needs switch of connected capillaries at ports 5 and 2. Also column inlet connections needed to be switched with outlet connections.
Page 128
Low dispersion heat exchanger or build-in 3 µL/6 µL heat exchanger and 90 mm length capillary to column Pos.1: Connection between ports 1-6 , 4-5, 2-3 , active column 1 = left (enrichment column) Pos. 2: Connection between ports 1-2, 3-4, 5-6 active column 2 = right (analytical colum) 1290 Infinity TCC User Manual...
Page 129
Low dispersion heat exchanger or build-in 3 µL/6 µL heat exchanger and 90 mm length capillary to column Pos.1: Connection between ports 1-6 , 4-5, 2-3 , active column 1 = left (Pre-column) Pos. 2: Connection between Ports 1-2, 3-4, 5-6 active column 2 = right (analytical column) 1290 Infinity TCC User Manual...
Page 130
Pos.1: Connection between ports 1-10 , 2-3, 4-5, 6-7, 8-9 active column 1 = left / regemerating column = right Pos. 2: Connection between ports 1-2, 3-4, 5-6, 7-8, 9-10 active column 2 = right / regenerating column = left 1290 Infinity TCC User Manual...
132 may vary. 1 Remove the front cover. 2 Use a pipette and tissue to dry the leak sensor area. 3 Observe the capillary connections and the column switching valve for leaks and correct, if required. 1290 Infinity TCC User Manual...
Page 132
Maintenance Correcting Leaks 4 Re-install the front cover. Figure 36 Possible Leak Areas 1290 Infinity TCC User Manual...
Stator face assembly Stator ring (available for service only) Rotor seal Bearing ring (P/N 1534-4045) Spanner nut (P/N 5068-0106, available for service only) When If valve leaks. Tools required Description Wrench, 1/4 inch Hexagonal key, 9/64 inch 1290 Infinity TCC User Manual...
Page 134
5 Remove the rotor seal (and isolation seal if damaged or contaminated). 6 Install the new isolation seal (if required). Ensure the metal spring inside the ring faces towards the valve body. 7 Install the new rotor seal. 1290 Infinity TCC User Manual...
Page 135
For example do not use 400 bar valve in a TCC or Flex Cube in combination with a 600 bar pump. 12 Perform a Pressure Test to ensure the valve is pressure tight. 1290 Infinity TCC User Manual...
When using multiple column compartments as part of a method development solution, make sure that the high pressure valve head is connected to the autosampler and the low pressure valve head is connected to the detector. 1290 Infinity TCC User Manual...
Page 137
Selection of valve port positions can fail, if the instrument does not know the properties of the installed valve. To have the valve correctly recognized by the Agilent Infinity Valve Drive you must have the N O T E valve drive powered off for at least 10 seconds. 1290 Infinity TCC User Manual...
Page 138
Screw the valve head onto the valve drive using the union lobe fits to the groove (see also “Installing the Valve nut (see also “Installing the Valve Heads” on page 50. Heads” on page 50). 1290 Infinity TCC User Manual...
Page 139
Maintenance Replacing Valve Heads Install all required capillary connections to the valve Push the valve head until it snaps in and stays in the rear head. position. Switch on the module. 1290 Infinity TCC User Manual...
Transportation Lock Kit (G1316-67001 ) can be re-ordered. Damage to Internal Parts C A U T I O N Mechanical shocks for example when being transported by car or shipped by post. ➔ Install a lock (Transportation Lock Kit). 1290 Infinity TCC User Manual...
Page 141
Preparing the Column Compartment for Transportation Remove the valve head as described in “Replacing Valve Replace the valve head by the transportation valve head. Heads” on page 136. Bring the transportation valve head to the outer position. 1290 Infinity TCC User Manual...
Page 142
Lock plate (G1316-03701) to the heat exchangers in the the Transportation valve head (G1316-40002) in the outer positions of the plate. central position of the plate. Put the spring washer (not shown) between the screw and the lock plate. 1290 Infinity TCC User Manual...
Initial firmware A.06.10 (main and resident) Compatible with All other modules must have firmware revision A.06.1x or B.06.1x or 1100/1200 series above (main and resident). Otherwise the communication will not work. modules Conversion to / emulation 1290 Infinity TCC User Manual...
Page 144
Maintenance Replacing the Module Firmware 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Parts and Materials for Maintenance Plastic Parts Leak Parts Heater and Cooling Devices Valve Options Overview Accessory Kits Standard Accessory Kit Low Dispersion Capillary Kit for G1316C Accessories This chapter provides information on parts for maintenance.
Parts and Materials for Maintenance Plastic Parts Plastic Parts Item Description G1316-68754 Front cover 5042-9964 Name plate for Agilent 1290 series Figure 38 Plastic Parts 1290 Infinity TCC User Manual...
Parts and Materials for Maintenance Leak Parts Leak Parts Item Description G1316-67000 Leak tube kit includes following items: Funnel holder G1316C, tubing-flex polyethylene, leak funnel G1316-42303 Capillary guide Figure 39 Leak Parts 1290 Infinity TCC User Manual...
Heater long-down ( 0.12 mm i.d., 1.6 µL internal volume) (G1316-80003) Part of Capillary kit (5067-4646). For fittings see Table 23 on page 150. Heat Exchanger Long Down 0.075 mm (nom. 1.0 µL) (G1316-80013) Column clip set, 8 colors (5042-9918) 1290 Infinity TCC User Manual...
Page 149
Heater and Cooling Devices Table 22 Heater Devices Item Description Carrier for heat exchanger TCC SL Plus (G1316-89200) Fitting holder assembly (G1316-68706) Bio-Inert Solvent Heating Devices ( 9 µL internal volume, 600 bar max pressure) (G5616-60050) 1290 Infinity TCC User Manual...
Page 150
Table 23 TCC Heater Fittings Item Description Plastic fittings (0100-1259) Very high pressure removable fitting (5067-4738) Fitting screw long Ferrule back Ferrule front included in kit Fitting screw long (5065-4454) Figure 40 TCC Heater Fittings 1290 Infinity TCC User Manual...
Page 151
Parts and Materials for Maintenance Heater and Cooling Devices Figure 41 Heater Devices for G1316C 1290 Infinity TCC User Manual...
Valve Head Parts (example shows a 5067-4159 valve head) Stator screws Stator head assembly Stator face assembly Stator ring (available for service only) Rotor seal Bearing ring (P/N 1534-4045) Spanner nut (P/N 5068-0106, available for service only) 1290 Infinity TCC User Manual...
Page 153
2 Pos/10 Port, 1200 bar (Vespel) 5067-4145 0101-1415 5068-0165 5068-0019 2 Pos/10Port, 600 bar (PEEK) 5067-4146 5068-0076 5068-0077 5068-0089 6 Column Selector, 600 bar (PEEK) 5067-4142 5068-0067 5067-0077 5068-0089 6 Column Selector, 1200 bar (Vespel) 1290 Infinity TCC User Manual...
CAN cable, Agilent module to module, 0.5 m G1316-01202 (2x) Column clamp G1316-87300 Capillary, 0.17 x 90 mm 1/16 in male/male G1316-87321 Capillary column-heat exchanger 105 mm lg, 0.17 mm i.d. G1316-87323 Capillary column-heat exchanger 170 mm lg, 0.17 mm i.d. 1290 Infinity TCC User Manual...
Page 160
Capillary ST 0.17 mm x 120 mm SX/SX valve to valve (bypass) 5067-4721 Capillary ST 0.17 mm x 200 mm S/SX valve to detector 0890-1713 Tube PTFE, 2 m valve to waste 5042-9918 Column clip set, eight colors 1290 Infinity TCC User Manual...
You will need this kit to prepare the TCC for transportation. Transportation Lock Kit (G1316-67001 ) Description G1316-03701 Lock plate 2680-0128 Screw M4 0515-0897 Screw M3x8 (pack of 4) Spring Washer G1316-40002 Transportation valve head 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Identifying Cables Cable Overview Analog cables Remote Cables BCD Cables CAN Cable RS-232 Cable Kit Agilent Module to Printer This chapter summarizes information on all cables. Agilent Technologies...
Identifying Cables Cable Overview Cable Overview Never use cables other than the ones supplied by Agilent Technologies to ensure proper N O T E functionality and compliance with safety or EMC regulations. Analog cables Description 35900-60750 Agilent module to 3394/6 integrators...
Page 165
It's also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9. 5181-1561 RS-232 cable, 8 m 1290 Infinity TCC User Manual...
Pin Agilent Signal Name module Not connected Shield Analog - Center Analog + Agilent Module to BNC Connector p/n 8120-1840 Pin BNC Pin Agilent Signal Name module Shield Shield Analog - Center Center Analog + 1290 Infinity TCC User Manual...
Page 167
Identifying Cables Analog cables Agilent Module to General Purpose p/n 01046-60105 Pin Agilent Signal Name module Not connected Black Analog - Analog + 1290 Infinity TCC User Manual...
Identifying Cables Remote Cables Remote Cables One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent modules. The other end depends on the instrument to be connected to. Agilent Module to 3396A Integrators...
Page 169
3 - Gray Start 4 - Blue Shut down 5 - Pink connected 6 - Yellow Power on High 7 - Red Ready High 8 - Green Stop 9 - Black Start request 13, 15 connected 1290 Infinity TCC User Manual...
Page 170
01046-60201 Wire Color Pin Agilent Signal Name Active module (TTL) White Digital ground Brown Prepare run Gray Start Blue Shut down Pink connected Yellow Power on High Ready High Green Stop Black Start request 1290 Infinity TCC User Manual...
CAN cable, Agilent module to module, 1 m LAN Cables Description 5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection) 5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection) 1290 Infinity TCC User Manual...
It's also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9. 5181-1561 RS-232 cable, 8 m 1290 Infinity TCC User Manual...
Agilent Module to Printer Agilent Module to Printer Description 5181-1529 Cable Printer Serial & Parallel, is a SUB-D 9 pin female vs. Centronics connector on the other end (NOT FOR FW UPDATE). For use with G1323 Control Module. 1290 Infinity TCC User Manual...
Page 176
Identifying Cables Agilent Module to Printer 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Hardware Information Firmware Description Electrical Connections Serial Number Information Rear view of the module Interfaces Overview Interfaces Setting the 8-bit Configuration Switch (without On-board) LAN Communication Settings for RS-232C Special Settings Early Maintenance Feedback Instrument Layout This chapter describes the TCC in more detail on hardware and electronics.
• an instrument specific section, called main system Resident System This resident section of the firmware is identical for all Agilent 1100/1200/1220/1260/1290 series modules. Its properties are: • the complete communication capabilities (CAN, LAN and RS-232C) • memory management • ability to update the firmware of the 'main system'...
Page 179
Update of main system can be done in the resident system only. Update of the resident N O T E system can be done in the main system only. Main and resident firmware must be from the same set. Figure 43 Firmware Update Mechanism 1290 Infinity TCC User Manual...
Page 180
All these specific informations are described in the documentation provided with the firmware update tools. The firmware update tools, firmware and documentation are available from the Agilent web. http://www.chem.agilent.com/_layouts/agilent/downloadFirmware.aspx?whid=69761 • 1290 Infinity TCC User Manual...
There are no externally accessible fuses, because automatic electronic fuses are implemented in the power supply. Never use cables other than the ones supplied by Agilent Technologies to ensure proper N O T E functionality and compliance with safety or EMC regulations.
• CN = China Alphabetic character A-Z (used by manufacturing) Alpha-numeric code 0-9, A-Z, where each combination unambiguously denotes a module (there can be more than one code for the same module) 00000 Serial number 1290 Infinity TCC User Manual...
K1367E HiP ALS Clinical Ed. G1377A HiP micro ALS G2258A DL ALS G5664A Bio-inert FC-AS G5667A Bio-inert Autosampler G4226A ALS Detectors G1314B VWD VL G1314C VWD VL+ G1314E/F VWD K1314F Clinical Ed. G4212A/B DAD K4212B DAD Clinical Ed. 1290 Infinity TCC User Manual...
Page 185
• CAN connectors as interface to other modules • LAN connector as interface to the control software • RS-232C as interface to a computer • REMOTE connector as interface to other Agilent products • Analog output connector(s) for signal output 1290 Infinity TCC User Manual...
• 8 data bit with no parity and • one start bit and one stop bit are always used (not selectable). The RS-232C is designed as DCE (data communication equipment) with a 9-pin male SUB-D type connector. The pins are defined as: 1290 Infinity TCC User Manual...
Page 187
RS-232C Connection Table Direction Function Ground Figure 45 RS-232 Cable Analog Signal Output The analog signal output can be distributed to a recording device. For details refer to the description of the module’s main board. 1290 Infinity TCC User Manual...
Page 188
APG Remote The APG Remote connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features as common shut down, prepare, and so on. Remote control allows easy connection between single instruments or systems to ensure coordinated analysis with simple coupling requirements.
Page 189
START REQUEST (L) Request to start injection cycle (for example, by start key on any module). Receiver is the autosampler. Special Interfaces There is no special interface for this module. 1290 Infinity TCC User Manual...
Switch settings provide configuration parameters for serial communication protocol and instrument specific initialization procedures. With the introduction of the Agilent 1260 Infinity, all GPIB interfaces have been removed. N O T E The preferred communication is LAN. 1290 Infinity TCC User Manual...
Data Bits Parity Use the following tables for selecting the setting which you want to use for RS-232C communication. The number 0 means that the switch is down and 1 means that the switch is up. 1290 Infinity TCC User Manual...
Page 192
Switches Parity No Parity Odd Parity Even Parity One start bit and one stop bit are always used (not selectable). Per default, the module will turn into 19200 baud, 8 data bit with no parity. 1290 Infinity TCC User Manual...
Save your methods and data before executing a forced cold start. If you use the following switch settings and power the instrument up again, a forced cold start has been completed. Table 35 Forced Cold Start Settings (without on-board LAN) Mode Select TEST/BOOT 1290 Infinity TCC User Manual...
EMF limits, and then reset the EMF counters to zero. The next time the EMF counters exceed the new EMF limits, the EMF flag will be displayed, providing a reminder that maintenance needs to be scheduled. 1290 Infinity TCC User Manual...
• the plastic layers help cushion the electronic and mechanical parts from physical shock, and • the metal inner cabinet shields the internal electronics from electromagnetic interference and also helps to reduce or eliminate radio frequency emissions from the instrument itself. 1290 Infinity TCC User Manual...
Page 196
Hardware Information Instrument Layout 1290 Infinity TCC User Manual...
1290 Infinity TCC User Manual Appendix General Safety Information The Waste Electrical and Electronic Equipment (WEEE) Directive (2002-96-EC) Lithium Batteries Information Radio Interference Sound Emission Agilent Technologies on Internet This chapter provides addition information on safety, legal and web. Agilent Technologies...
C A U T I O N alerts you to situations that could cause loss of data, or damage of equipment. ➔ Do not proceed beyond a caution until you have fully understood and met the indicated conditions. 1290 Infinity TCC User Manual...
Page 199
The operator of this instrument is advised to use the equipment in a manner as specified in this manual. Safety Standards This is a Safety Class I instrument (provided with terminal for protective earthing) and has been manufactured and tested according to international safety standards. 1290 Infinity TCC User Manual...
Page 200
When working with solvents, observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet by the solvent vendor, especially when toxic or hazardous solvents are used. 1290 Infinity TCC User Manual...
With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a Monitoring and Control Instrumentation product. Do not dispose off in domestic household waste N O T E To return unwanted products, contact your local Agilent office, or see www.agilent.com more information. 1290 Infinity TCC User Manual...
Ved udskiftning benyttes kun batteri som anbefalt av apparatfabrikanten. ➔ Brukt batteri returneres appararleverandoren. Bij dit apparaat zijn batterijen geleverd. Wanneer deze leeg zijn, moet u ze niet weggooien N O T E maar inleveren als KCA. 1290 Infinity TCC User Manual...
Appendix Radio Interference Radio Interference Cables supplied by Agilent Technologies are screened to provide optimized protection against radio interference. All cables are in compliance with safety or EMC regulations. Test and Measurement If test and measurement equipment is operated with unscreened cables, or...
This product has a sound pressure emission (at the operator position) < 70 dB. • Sound Pressure Lp < 70 dB (A) • At Operator Position • Normal Operation • According to ISO 7779:1988/EN 27779/1991 (Type Test) 1290 Infinity TCC User Manual...
Appendix Agilent Technologies on Internet Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com 1290 Infinity TCC User Manual...
Page 207
CAN partner radio interference remote GLP features cable maintenance repairs feedback changing the column replacing firmware heater profile column switching valve parts materials heatexchanger correcting leaks bio-inert pre-column definition of message heating concept introduction remote timeout 1290 Infinity TCC User Manual...
Page 208
RFID access failure special interfaces verification special settings temperature boot-resident voltage range forced cold start specification physical warm-up time specifications waste status indicator electrical and electronic system overview equipment WEEE directive weight installation temperature calibration 1290 Infinity TCC User Manual...
Page 210
In This Book The manual contains technical reference information about the Agilent 1290 Infinity Thermostatted Column Compartment. This manual describes the following: • introduction and specifications, • installation, • using and optimizing, • troubleshooting and diagnose, • maintenance, • parts identification, •...