Invalid Motor Status Motor AUX1. (AUX2)
Probable cause
1
2
3
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
Injection valve not connected.
Defective LCVlv1 Connection cable.
Control Board defective.
LC Injector HTC/HTS Description and Installation
Suggested actions
Switch OFF LC Injector HTC/HTS and check
connection between LC Injector HTC/HTS and
Valve Drive.
Change connection cable.
Replace Control-HTC or Control-xt Board.