4 After selecting "Named Tray Holder" the X- Y- Z-positions for the object will be
5 Highlight item Position X with the cursor bar and press ENTER. The Injection
6 Rotate the outer knob to adjust the X-axis position to the Tray Holder
7 Press the inner knob to ENTER the Position X value.
8 Repeat steps 5-7 for Position Y and Position Z.
9 If F3 "Moveto Zero" is activated, the Injection Unit will move to the HOME
10 Verify the defined X-, Y-, Z-positions by pressing F1 "Check Pos".
Figure 40
are given below.
Figure 40
The release of Firmware level 4.1.X makes possible the correction of all three axes, X-,Y-, Z,
N O T E
in the firmware class Tray. The correction of the inclination can be made for a tilted Tray in
the direction of the row and/or the column. Teaching is possible in the Utilities section by
using the path:
Menu/Utilities/Trays.
These functions are not available for the LC Injector HTC/HTS if operated solely on PAL
N O T E
Firmware level 2.4.X or higher, but neither with firmware level 3.X.X. nor with level 4.X.X.
Agilent 1290 Infinity LC Injectors HTC/HTS User Manual
displayed (see
Figure 39
Unit will move to the previously defined X-axis position.
reference position.
position;
on page 89illustrates possibilities of correction. Details for teaching
Inclined Tray, Corrections for X-, Y-, Z-Axes
LC Injector HTC/HTS Description and Installation
on page 88).
3
Object Positions
89